TB 9-6625-086-35
10
(4) Set the TI MOUNT RES switch to agree with the thermistor mount (A8) being
used.
(5) Set the TI RANGE switch to .01 MW and null and zero the meter.
(6) Adjust dc power supply (A2) for a full-scale (.01 mw) indication on the TI. If dc
voltmeter (A3) indication is not within the limits specified in table 5, perform
b
below.
Figure 4. Range and tracking check without power meter calibrator.
Table 5. Range and tracking Performance Check (Without Power Meter Calibrator)
Test
Test
Dc Voltmeter Indications (vdc) (Readings
Adjustments for Model
Instrument
Instrument
for Model 431B are in Parenthesis)
Switch
Indication
200-Ohm Mount
100-Ohm Mount
200-Ohm
100-Ohm
Settings
Min
Max
Min
Max
Mount
Mount
.01
.01
.438
(.434)
.456
(.460)
.619
(.613)
.645
(.651)
R14
R1
.03
.03
.763
(.752)
.787
(.798)
1.08
(1.06)
1.12
(1.13)
R13
R2
.1
.1
1.40
(1.37)
1.42
(1.45)
1.98
(1.94)
2.02
(2.06)
R12
R3
.3
.3
2.43
(2.38)
2.47
(2.52)
3.43
(3.36)
3.49
(3.56)
R11
R4
1
1
4.43
(4.34)
4.51
(4.60)
6.26
(6.13)
6.38
(6.51)
R10
R5