5
A
djustment
Introduction
This
chapter
describes
the
adjustments
required
to
return
the
HP
41802A
to
a
condition
to
meet
the
specications
listed
in
Figure
1-2,
if
the
HP
41802A
failed
the
performance
test,
or
after
it
has
been
repaired.
If
proper
performance
cannot
be
achieved
after
these
adjustments
,
refer
to
Chapter
6.
Equipment
Required
T
able
4-1
lists
the
equipment
required
for
adjustment.
Preparation
This
paragraph
provides
the
information
which
you
should
know
,
and
the
procedures
that
you
must
perform
before
starting
the
adjustments
.
1.
Allow
the
HP
41802A
and
the
test
equipment
to
warm
up
and
stabilize
for
at
least
30
minutes
before
you
execute
the
adjustments
.
2.
P
erform
the
adjustments
in
an
ambient
temperature
of
23
C
6
5
C.
Adjustment
5-1