Maximum
Allowable
Level
Caution
Maximum
allowable
level
(ac+dc)
to
the
probe
is
650
V
.
Discharging
the
Probe
Caution
Measuring
a
node
having
a
dc
voltage
potential
charges
the
blocking
capacitors
inside
the
HP
41802A.
Ground
the
probe
pin
after
measuring
such
nodes
to
discharge
probe
capacitor
.
F
ailure
to
do
this
could
result
in
damage
to
sensitive
circuits
in
the
DUT
,
especially
if
it
is
an
active
device
.
Probe
P
ower
Plug
Note
Connect
the
probe
power
plug,
before
connecting
the
N-type
connector
.
Preparation
for
Use
This
paragraph
provides
the
HP
41802A
operating
check
procedure
and
the
probe
adjustment
procedure
.
The
HP
41802A
operating
check
procedure
checks
the
adapter
gain
using
a
network
analyzer
,
a
spectrum
analyzer
,
or
a
network/spectrum
analyzer
.
These
operating
check
procedures
are
only
intended
to
ensure
that
the
HP
41802A
is
functional.
If
the
HP
41802A
fails
this
check,
it
should
be
repaired.
T
o
verify
that
the
HP
41802A
meets
its
specications
,
perform
the
performance
test
described
in
Chapter
4.
The
probe
adjustment
procedure
is
for
frequency
compensation.
F
or
the
best
measurement
results
,
always
adjust
the
probe
.
Operation
3-3