TM 11-6625-1614-15
.
Model 410C
Parsgraphs 5-20 to 5-23
Figure 5-3.
High Frequency Response Test
c. Set Model 410C FUNCTION SELECTOR to
+DCV ; RANGE to .015 V.
d. Adjust voltmeter calibrator for +. 015 VDC
output.
e. The dc voltmeter should read +1.5 v. This
will vertfy a gain of 100, when the gain /A/
equals
E
DC out/
E
410C.
5-20. AMPLIFTER NOfSE CHECK.
a. Leave the dc voltmeter connected to the DC
AMPLIFIER OUTPUT as in Paragraph 5-19.
b. Set the Model 410C RANGE to 1500 V;
FUNCTION SELECTOR to +DCV.
c. Short the Model 410C DCV and COM cables.
Note dc voltmeter readings.
This
reading
s h o u l d b e l e s s t h a n 7 . 5 m i l l i v o l t s .
d. Reset Model 410C RANGE to 1.5 V. DC
Voltmeter should read less than 7.5 mv.
5-21.
DC AMPLIFTER OUTPUT
IMPEDANCE
CHECK.
a. Connect an external DC Voltmeter Model
3440A/3442A) to Model 410C DC AMPLIFIER
OUTPUT terminals on rear panel.
b. Set Model 410C FUNCTION SELECTOR to
OHMS position.
c . R e c o r d v o l t a g e i n d i c a t e d o n e x t e r n a l d c
voltmeter for use as a reference.
d. Connect a 1.5 k ohm ±1% resistor Part
No. 0730-0017) across Model 410C DC AMPLI -
FIER OUTPUT terminals. DC voltage recorded
in step c above should not change more than
3 mv, indicating that dc amplifier output im-
pedance is within the 3 ohm specification at dc.
5-22. AC VOLTMETER OPERATION.
5-23. 11036A
AC
PROBE ACCURACY CHECK.
a. Figure 5-3 describes the test arrangement
required for this operation. Do not
place
Model 410C AC Probe in T-Connector at this
point.
b. Adjust signal generator for a 0.7 volt (rms)
output at 1000 cps.
c. Connect Model 11036A AC Probe to signal
generator and read output on Model 410C Volt-
meter (meter should read 0.7 volts).
d. Remove probe tip from Model 11036A and
connect the ac probe as shown in Figure 5-4.
e. Turn signal generator to 50 Mc and adjust
signal generator for a power reading of 9.8
dbm (0.7 volts) on the power meter.
f. The difference between reading on Model
410C meter and 0.7 volt reference is the ac
probe error at that frequency.
g. Repeat steps f and g every 100 Mc from 50
to 700 Mc.
01556-2
5-5
Summary of Contents for 410C
Page 6: ...Figure 1 1 TM 11 6625 1614 15 M o d e l 4 1 0 C Section I Figure 1 1 1 0 ...
Page 8: ......
Page 9: ...Table 1 1 TM 11 6625 1614 15 Section I Table 1 1 Model 410C 01556 2 1 3 ...
Page 22: ...Figure 3 5 TM 11 6625 1614 15 Section III Figure 3 5 Model 410C 3 8 0 1 5 5 6 2 ...
Page 29: ...Figure 4 3 Model 410C TM 11 6625 1614 15 Section IV Figure 4 3 01556 2 4 3 ...
Page 30: ...Figure 4 4 TM 11 6625 1614 15 Section IV Figure 4 4 Model 410C 4 4 01556 2 ...
Page 31: ...Figure 4 5 TM 11 6625 1614 15 Section IV Figure 4 5 01556 2 4 5 ...
Page 32: ...Figure 4 6 TM 11 6625 1614 15 Section IV Figure 4 6 Model 410C 4 6 01556 2 ...
Page 33: ......
Page 44: ......
Page 47: ...Table 5 8 M o d e l 4 1 0 C TM 11 6625 1614 15 5 1 3 ...
Page 53: ...Figure 5 6 1 TM 11 6626 1614 15 Change 1 5 16 1 ...
Page 54: ......
Page 55: ...TM 11 6625 1614 15 Section V Figure 5 8 Figure 5 8 Power Supply Schematic 5 19 ...
Page 56: ...5 20 TM 11 6625 1614 15 Figure 5 9 Typical Amplifier Waveforms ...
Page 57: ......
Page 58: ......
Page 59: ......
Page 61: ......
Page 62: ...THE METRIC SYSTEM AND EQUIVALENTS ...
Page 63: ...PIN 016288 000 ...