HP 410C Organizational, Ds, Gs, And Depot Maintenance Manual Download Page 27

TM 11-6625-1614-15

Model 410C

Section IV

Paragraph 4-1 to 4-18

4-1.

4-2.

OVERALL DESCRIPTION.

The Model 410C includes

SECTION IV

THEORY OF OPERATION

an input network, a

modulator - amplifier- demodulator, and a

meter

circuit. A block diagram of the Model 410C is shown
in Figure 4-1.

4-3. Signals to be measured are applied through the
appropriate input lead to the input network. AC volt-
ages are detected in the AC probe, and therefore all
signals to the input network are DC. The input net-
work attenuates the DC signal to a level determined
by RANGE and FUNCTION SELECTOR settings. The
attenuated DC voltage is applied to the modulator which
converts the DC to AC for amplification. The ampli-
fied AC signal is converted back to DC voltage inthe
demodulator and coupled to cathode follower VIB. The
cathode follower output to the DC AMPLIFIER OUT-
PUT connector and meter circuit is a DC voltage
proportional to the amplitude of the signal applied to
the input. A portion of the voltage to the meter circuit
is returned to the modulator as feedback. When the
feedback voltage and attenuated DC voltage are nearly
equal, the meter stabilizes.

4-4. CIRCUIT DESCRIPTION.

4-5. INPUT NETWORK.

4-6. The input network includes a precision voltage
divider, which by means of the FUNCTION SELECTOR
and RANGE switches, providesa maximum of 15 milli-
volts at the modulator input regardless of the range
set and signal applied. The ± DCA, ±DCV, OHMS, and
ACV modes of operation are discussed below.

4-7. DC CURRENT MEASUREMENTS: Refer to Fig-
ure 4-3, throughout this explanation. The purpose of
the input network is to provide proper attenuation of
currents applied.

Currents from 1.5 µa to 150 ma

full scale are applied with input impedance decreasing
from 9K ohms on the 1.5 µa range to approximately
0.3 ohms on the 150 ma range.

4-6. Tbe change in input impedance is varied by using
DC current shunts in conjunction with RANGE switch
A2S2. The DC voltage developed across these shunt

resistors, when applied through the modulator-am-

plifier-demodulator network to the meter, provide a
deflection on the meter proportional to the DC current
being measured.

4-9. DC VOLTAGE MEASUREMENTS.

R e f e r   t o

Figure 4-4 throughout this explanation. The purpose
of the input network is to accurately attenuate the in-

put signal to a maximum of 15 millivolts at the modu-

01556-2

later input. The network presents an input impedance
of 10 megohms on the three most sensitive ranges and
100 megohms on all other ranges.

4-10. The resistor R1 (located in the DCV probe) in
conjunction with resistors A2R10 through A2R26, pro-
vides the 10 megohm input impedance required for the
three most sensitive DCV ranges. Resistors A2R4

and A3R30 are shunted out of the circuit by the RANGE

switch on the three most sensitive DCV ranges.

4-11. When using the eight less sensitive ranges,
A2R4 and A3R30 are placed in series with Rl and

A2R10 through A2R26 to present more than 100 meg-

ohm impedance to the input.

4-12. A3R30 is used to calibrate full scale on the

1500 volt range. (See Paragraph 5-35. )

4-13. RESISTANCE MEASUREMENTS. The purpose
of the input network shown in Figure 4-5 is to place

approximately 0. 6 volt DC source in series with a
known (reference) resistance. The resistance to be

measured is ptaced in parallel with the known resis-
tance, which changes the voltage proportionally. The
maximum changes in voltage applied to the modulator
is 15 mv because of attenuation provided by A2R4,
A3R30, and A1R2.

4-14.

A DC current of approximately 60 ma is

supplied at the junction of A2R22 and A2R23 through

A7R10, R2, A2R2 and A2R1 to the input network. The
OHMS   ADJ., R3, sets the meter for full scale

Resistor A2R1 is shorted out in the XIM position of

the RANGE switch; resistors A2R1 and A2R2 are
shorted out in the X10M range. The resistors A2R2
and/or A2R1 are electrically removed from the cir-
cuit to increase the voltage at the junction of A2R22
and A2R23. This is done to compensate for tbe load-
ing of the attenuator (A2R4, A3R30, and A1R2) on
these ranges.

4-15.

AC VOLTAGE MEASUREMENTS. Refer to

Figure 4-6 throughout this explanation. Voltage at
the AC probe is converted to DC and applied to the in-

put network. The input signal is attenuated to produce
a maximum of about 15 millivolts at the modulator in-
put .

AC zero adjustment of meter pointer is made

with the AC ‘ZERO control.

4-16. MODULATOR-DEMODULATOR.

4-17. Refer to the Amplifier Schematic, Figure 5-10 ,
and to the Mechanical Analogy Schematic,

Figure

4-2 throughout this explanation.

4-18. The input network applies approximately 15
millivolts DC, for full scale meter deflection (posi-
tive or negative, depending on the polarity of the

4-1

Summary of Contents for 410C

Page 1: ... A L ORGANIZATIONAL DS GS AND DEPOT MAINTENANCE MANUAL HEWLETT PACKARD ELECTRONIC VOLTMETER MODEL 410C T h i s c o p y I S a r e p r i n t w h i c h i n c l u d e s c u r r e n t p a g e s f r o m C h a n g e 1 H E A D Q U A R T E R S D E P A R T M E N T O F T H E A R M Y AUGUST 1967 ...

Page 2: ...TM 11 6625 1614 15 WARNING DANGEROUS VOLTAGES EXIST IN THIS EQUIPMENT Be careful when working on the power supplies and their circuits or on the 115 or 230 volt ac line connections DON T TAKE CHANCES ...

Page 3: ... Conductor Power Cable 2 12 Primary Power Requirements 2 14 Repackaging for Shipment III OPERATION 3 1 Introduction 3 3 Adjustment of Mechanical Zero 3 5 Front and Rear Panel Description 3 7 Operating Procedures 3 9 DC Voltage Measurements Figure 3 2 3 11 DC Current Measurements Figure 3 3 3 13 AC Voltage Measurements Figure 3 4 3 15 Precautions When Measuring AC Voltage 3 28 Negative Pulses 3 31 ...

Page 4: ...age of Complex Waveforms Recommended Test Equipment DCV Accuracy Test DCV Input Resistance Test DCA Accuracy Test Deleted AC Accuracy Test Power Supply Test AC Full Scale Adjust Front Panel Troubleshooting Prmedure Troubleshooting Prwedure LIST OF ILLUSTRATIONS The Combining Case Steps to Place Instrument in Combining Case Adaptor Frame Instrument Combination TWO Half Modules in Rack Adaptor Front...

Page 5: ...ion High Frequency Response Test Low Frequency Response Test Troubleshooting Tree A4 Chopper Assembly Installation Chopper Frequency Adjust Setup Power Supply Measurements Power Supply Schematic Typical Amplifier Waveforms Amplifier Schematic Model 11036A AC Probe Exploded Model l1036A Probe Schematic RANGE and FUNCTION Switching Pictorial Input RANGE and FUNCTION Switching Schematic Page 4 4 4 5 ...

Page 6: ...Figure 1 1 TM 11 6625 1614 15 M o d e l 4 1 0 C Section I Figure 1 1 1 0 ...

Page 7: ...5 73 AFR 400 54 MCO 4430 3E c Discrepancy in Shipment Report DISREP SF361 Fill out and forward Discrepancy in Ship ment Report DISREP SF 361 as prescribed in AR 55 38 NAVSUPINST 4610 33B AFR 75 18 MCO P4610 19C and DLAR 4500 15 1 A 4 Reporting Errors and Recoin mending Improvements You can help improve this manual If you find any mistakes or if you know of a way to improve the pro cedures please l...

Page 8: ......

Page 9: ...Table 1 1 TM 11 6625 1614 15 Section I Table 1 1 Model 410C 01556 2 1 3 ...

Page 10: ...e can be measured with an accuracy of 5 of reading at midscale resistance from 0 2 ohms to 500 megohms can be measured with reduced accuracy The Model 410C Electronic Voltmeter is shown in Figure 1 1 the specifications are given in Table 1 1 1 3 With the Model 11036A detachable AC Probe the Voltmeter can be used to measure AC voltage from 20 cps to 700 Mc From 20 cps to 100 MC AC voltage from 0 5 ...

Page 11: ...over a frequency range of 20 cps to 700 Mc Reference calibration accuracy at 400 cps sinusoidal is 3 of full scale Frequency response is 10 from 20 cps to 700 Mc with indications obtainable to 3000 Mc Frequency response at 100 Mc is within 2 The Model 110364 responds to the positive peak above average value of the signal applied The Model 410C is calibrated to read in RMS volts for sine wave input...

Page 12: ...Model 410C TM 11 6625 1614 15 Section II Figures 2 l and 2 2 Figure 2 1 The Combining Case Figure 2 2 Steps to Place Instrument in Combining Case 2 0 01556 1 ...

Page 13: ...ombinations of submodular units It can be rack mounted only An illustration of the adapter frame is given in Figure 2 3 Instruc tions are given below a Place the adapter frame on edge of bench as shown in step 1 Figure 2 4 b Stack the submodular units in the frame as shown instep 2 Figure 2 4 Place the spacer clamps between instruments as shown in step 3 Figure 2 4 c Place spacer clamps on the two...

Page 14: ...s a Place instrument in original container if avai Model 410C lable Figure 2 4 Two Half Modules in Rack Adapter b Ensure that container is well sealed with strong tape or metal bands 2 17 If original container is not to be used proceed c Place instrument and inner container in a heavy as follows carton or wooden box and seal with strong tape or a Wrap instrument in heavy paper or metal bands plast...

Page 15: ...3 3 3 12 General instructions for the measurement of DC current are the same as those given for DC volt age measurements Paragraph 3 9 01556 2 3 13 1 3 14 AC VOLTAGE MEASUREMENT Figure 3 4 ONE SIDE OF ALMOST ALL POWER DISTRIBUTION SYSTEMS IS GROUNDED EXTREME CAUTION MUST BE USED IF DIRECT MEASUREMENT OF POWER LINE VOLTAGES IS ATTEMPTED IF THE GROUND CLIP LEAD IS ACCIDEN TALLY CONNECTED TO THE UN G...

Page 16: ...r can be con nected from the floating ground terminal to the chassis ground terminal on the DC AMPLIFIER OUTPUT connector If a shorting bar is not used the COM Lead is floating except when the FUNCTION SELECTOR is set to ACV 8 9 10 11 12 13 14 DCV This lead is used in conjunction with the COM Lead to measure DC voltage AC PROBE 300V MAX Receptacle for tele phone type plug of Model 11036A AC Probe ...

Page 17: ...isting coaxial transmission line cut the line away so the center con ductor of the line is exposed through a hole large enough to clear the body of the probe The nose of the probe should be removed for this type of measurement Connect one terminal of a button type capacitor of ap proximately 50 picofarads to the center conductor of the coaxial line so that the other terminal of the oapa citor will...

Page 18: ...0 pps t l would be 990 microseconds Thus To t 1 would be approximately 0 and from the graph it is seen that K is approximately 0 The expression would then reduce to 3 30 It can be seen that in the case of negative pulses of short duration much smaller readings will be ob tained for an equivalent positive pulse As a result large multiplying factors must be used and unless the pulse voltage is large...

Page 19: ...TM 11 6625 1614 15 Model 410C Section III F i g u r e 3 2 01556 2 Figure 3 2 DC Voltage Measurements 3 5 ...

Page 20: ...TM 11 6625 1614 15 Section III Model 410C Figure 3 3 Figure 3 3 DC Current Measurements 3 6 01556 3 ...

Page 21: ...TM 11 6625 1614 15 Model 410C Section III Model 410C Figure 3 4 AC Voltage Measurements 01556 3 3 7 ...

Page 22: ...Figure 3 5 TM 11 6625 1614 15 Section III Figure 3 5 Model 410C 3 8 0 1 5 5 6 2 ...

Page 23: ...TM 11 6625 1614 15 Model 410C Section III Figure 3 6 Figure 3 6 Graph Used in Calculation of Pulse Voltage Readings 01556 2 3 9 ...

Page 24: ...TM 11 6625 1614 15 Section III Model 410C Figure 3 7 Figure 3 7 Resistance Measurements 3 10 01556 2 ...

Page 25: ...TM 11 6625 1614 15 Model 410C Section III Figure 3 8 Figure 3 8 DC Nano Amoere Current Measurements 01556 3 3 11 ...

Page 26: ...TM 11 6625 1614 15 Section IV Figure 4 1 and 4 2 Model 410C Figure 4 1 Block Diagram Model 410C Figure 4 2 Modulator Demodulator Mechanical Analogy 4 0 01556 2 ...

Page 27: ...f the input network is to accurately attenuate the in put signal to a maximum of 15 millivolts at the modu 01556 2 later input The network presents an input impedance of 10 megohms on the three most sensitive ranges and 100 megohms on all other ranges 4 10 The resistor R1 located in the DCV probe in conjunction with resistors A2R10 through A2R26 pro vides the 10 megohm input impedance required for...

Page 28: ...RK 4 24 The feedback network drives the meter and determines the DC gain of the amplifier The feed back is varied depending on the position of the FUNC TION and RANGE selectors The different feedback configurations are discussed below 4 25 FEEDBACK NETWORK FOR DCA OHMS AND DCV Figures 4 3 4 4 and 4 5 show the feed back configuration for ail positions of the FUNCTION SELECTOR except ACV The meter i...

Page 29: ...Figure 4 3 Model 410C TM 11 6625 1614 15 Section IV Figure 4 3 01556 2 4 3 ...

Page 30: ...Figure 4 4 TM 11 6625 1614 15 Section IV Figure 4 4 Model 410C 4 4 01556 2 ...

Page 31: ...Figure 4 5 TM 11 6625 1614 15 Section IV Figure 4 5 01556 2 4 5 ...

Page 32: ...Figure 4 6 TM 11 6625 1614 15 Section IV Figure 4 6 Model 410C 4 6 01556 2 ...

Page 33: ......

Page 34: ... Measure Model 3400A RMS ments ripple Voltmeter Electronic Counter Frequency Range to at Chopper Frequency Adjust Model 521C Electronic least 102 cps Counter DC Standard output 1000 v DC Adjust Model 740A DC Accuracy 0 2 Standard Ohmmeter Range 100 Mf2 Troubleshooting Model 412A DC VTVM Accuracy 5 Thermistor Mount Frequency 10 Mc to 700Mc Frequency Response Test Model 478A Coaxial Impedance 50 ohm...

Page 35: ...nducted 5 7 ALTERNATE VOLTAGE SOURCE 5 6 Should it be necessary to use the Voltmeter Calibrator to conduct these Performance Checks the arrangement described in Figure 5 1 will provide the necessary voltage values required How ever the Model 738BR Voltmeter Calibrator is the 5 9 Mechanical METER ZERO a Turn instrument on Allow at least a 20 minute warm up period b Turn voltmeter off and allow 30 s...

Page 36: ... Rin on RANGES specified 5 13 5 2 Voltage 015 05 15 5 1 5 5 15 50 150 300 300 Model 410C Meter Readings 01 47 to 0153 V 049 to 051 V 147 to 153 V 49 to 51 V 1 47 to l 53 V 4 9 to 5 l V 14 7 to 15 3 V 49 to 51 V 147 to 153 V 290 to 310 V 270 to 330 V 5 14 ACCURACY CHECK DCA a Figure 5 2 describes the test arrangement required for this operation The following additional equipment will also be requir...

Page 37: ...TM 11 6625 1614 15 Model 410C Section V Tables 5 3 and 5 4 Figure 5 2 Figure 5 2 DC Ammeter Operation Table 5 3 DCV Input Resistance Test Table 5 4 DCA Accuracy Test 01556 2 5 3 ...

Page 38: ... FUNCTION SELECTOR to OHMS RANGE to RX10 c Set pointer to using rear panel adjust ment OHMS ADJ if required h If both of these ranges function properly it can be assumed that the remainder will also If meter does not function properly refer to Paragraph 5 31 for adjustment procedure 5 17 AMPLIFIER OPERATION Deleted see paras 5 19 and 5 24 d Connect COM and DCA OHMS cables across 10 ohm resistor 5 ...

Page 39: ... Record voltage indicated on external dc voltmeter for use as a reference d Connect a 1 5 k ohm 1 resistor Part No 0730 0017 across Model 410C DC AMPLI FIER OUTPUT terminals DC voltage recorded in step c above should not change more than 3 mv indicating that dc amplifier output im pedance is within the 3 ohm specification at dc 5 22 AC VOLTMETER OPERATION 5 23 11036A AC PROBE ACCURACY CHECK a Figu...

Page 40: ...librator to settings listed in Table 5 6 Model 410C should indicate readings within limits specified If not refer 5 25 AC VOLTMETER FREQUENCY RESPONSE CHECK a A Frequency Response Test Set Model 739AR a Test Oscillator Model 651A an RF Signal Generator Model 608 C a Power Meter Model 431 B a Thermistor Mount Model 478A a Probe T Connector Model 11042A a VHF Signal Generator Model 612A and a 10 KC ...

Page 41: ... r o b e t i p o n t h e M o d e l 1 1 0 3 6 A a n d c o n n e c t d i r e c t l y t h r o u g h a 5 0 o h m l o a d t o t h e o u t p u t o f t h e F r e q u e n c y R e s p o n s e T e s t S e t C o n n e c t t h e o u t p u t o f t h e T e s t O s c i l l a t o r d i r e c t l y t o t h e i n p u t o f t h e F r e q u e n c y R e s p o n s e T e s t S e t O b s e r v e s t e p g t h r o u g h o...

Page 42: ... l i b r a t i o n p r o c e d u r E f o r t h e M o d e l 4 1 0 C T h e s e o p e r a t i o n s s h o u l d b e c o n d u c t e d o n l y i f i t h a s p r e v i o u s l y b e e n e s t a b l i s h e d b y P e r f o r m a n c e C h e c k s P a r a g r a p h 5 5 t h a t t h e M o d e l 4 1 0 C i s o u t o f a d j u s t m e n t I n d i s c r i m i n a t e a d j u s t m e n t o f t h e i n t e r n a...

Page 43: ...ple should not exceed 2 5 mv 5 3 0 D C Z e r o A d j u s t m e n t a n d B i a s a Set Model 410C Function Selector to DCV and Range Switch to 5 V b S h o r t D C V C a b l e t o C O M C a b l e c A d j u s t A 3 R 2 1 f u l l y c o u n t e r c l o c k w i s e a n d then rotate about 20 clockwise d Adjust ZERO ADJ pot on rear panel for zero meter deflection Switch to DCV If any deflection is obser...

Page 44: ......

Page 45: ... h e a b o v e r a n g e s r e a d j u s t A 6 R 1 8 t o r e d u c e e r r o r g I f e r r o r g r e a t e r t h a n 2 o f f u l l s c a l e e x i s t s o n a n y r a n g e b e t w e e n 1 5 m v a n d 1 5 0 m v i n c l u s i v e s e l e c t n e w s e t t i n g f o r A 6 R 1 8 t o y i e l d b e s t r e s u l t s o n t h e s e r a n g e s I f e r r o r g r e a t e r t h a n 2 o f f u l l s c a l e s...

Page 46: ... m o n s h o r t l e a d d Adjust R1 for Model 410C zero deflection 5 35 AC FULL SCALE ADJUST 5 V RANGE a C o n n e c t M o d e l 4 1 0 C a c p r o b e t o v o l t m e t e r c a l i b r a t o r o u t p u t t e r m i n a l s Set Model 410C FUNCTION SELECTOR to ACV RANGE to 500 v b A d j u s t v o l t m e t e r c a l i b r a t o r t o p r o v i d e 3 0 0 v 4 0 0 c p s r m s output M o d e l 4 1 0 C ...

Page 47: ...Table 5 8 M o d e l 4 1 0 C TM 11 6625 1614 15 5 1 3 ...

Page 48: ... isolating malfunctions The steps in Table 5 10 des cribe the normal conditions which should be encoun tered during the checks circled numbers in Figure 5 5 Figure 5 5 5 41 The checks outlined in Table 5 10 are not de signed to measure all circuit parameters rather only to localize the malfunction Therefore it is quite possible that additional measurements will be required to completely isolate th...

Page 49: ... e Clean excess flux from the connection and adjoining area a Use a low heat 25 to 50 watts small tip f To avoid surface contamination of the printed soldering iron and a small diameter rosin circuit clean with weak solution of warm water core aoider and mild detergent after repair Rinse thoroughly with clean water When completely dry spray b Circuit components can be removed by lightly with Krylo...

Page 50: ...Table 5 9 Paragraph 5 34 Table 5 10 TM 11 6625 1614 15 Model 410C Section V Table 5 9 01566 2 5 1 6 ...

Page 51: ...Figure 5 7 Table 5 10 Paragraph 5 29 Figure 5 10 Figure 5 7 Table 5 9 Figure 5 10 Figure 5 10 Table 5 9 TM 11 6625 1614 15 Section V Model 410C Table 5 10 5 1 7 ...

Page 52: ...TM 11 6625 1614 15 Model 410C Figure 5 11 Model 11036A AC Probe Exploded View 5 22 Figure 5 12 Model 11036A AC Probe Schematic 01556 2 ...

Page 53: ...Figure 5 6 1 TM 11 6626 1614 15 Change 1 5 16 1 ...

Page 54: ......

Page 55: ...TM 11 6625 1614 15 Section V Figure 5 8 Figure 5 8 Power Supply Schematic 5 19 ...

Page 56: ...5 20 TM 11 6625 1614 15 Figure 5 9 Typical Amplifier Waveforms ...

Page 57: ......

Page 58: ......

Page 59: ......

Page 60: ...l United States Army The Adjutant General Distribution Active Army USAMB 1 USACDCEC 1 USACDCCEA 1 USA CDCCEA Ft Huachuca 1 NG None USAR None For explanation of abbreviations used see AR 320 50 HAROLD K JOHNSON General United States Army Chief of Staff Eighth USA 5 SAAD 5 TOAD 5 LEAD 3 ...

Page 61: ......

Page 62: ...THE METRIC SYSTEM AND EQUIVALENTS ...

Page 63: ...PIN 016288 000 ...

Page 64: ...e it They did not even scan it Why should they get your money Why are not letting you give those free manuals to your friends I am setting this document FREE This document was made by the US Government and is NOT protected by Copyright Feel free to share republish sell and so on I am not asking you for donations fees or handouts If you can please provide a link to liberatedmanuals com so that free...

Reviews: