TM 11-6625-1614-15
Model 410C
5 - 1 . I N T R O D U C T I O N .
SECTION V
MAINTENANCE
5-2.
This section contains maintenance procedures
for the Model 410C Electronic Voltmeter.
5-3. TEST EQUIPMENT REQUIRED.
5-4. The test equipment required to maintain and
adjust the Model 410C is listed in Table 5-1. Equi-
pment having similar characteristics may be substi-
tuted for items listed.
5-5. PERFORMANCE CHECKS.
5 - 6 . T h e p e r f o r m a n c e c h e c k s p r e s e n t e d i n t h i s
section are front panel operations designed to com-
pare the Model 410C with it’s published specifications.
These operations may be incorporated in periodic
maintenance, post repair and incoming quality control
checks. These operations should be conducted before
any attempt is made at instrument calibration or
adjustment. During performance checks, periodically
vary the line voltage to the Model 410C, ± 10% on either
115v or 230 v operation. A 1/2 hour warm-up period
should be allowed before these tests are conducted.
5-7. ALTERNATE VOLTAGE SOURCE.
5-6. Should it be necessary to use the
Voltmeter Calibrator to conduct these Performance
Checks, the arrangement described in Figure 5-1 will
provide the necessary voltage values required. How-
ever; the Model 738BR Voltmeter Calibrator is the
5 - 9 . M e c h a n i c a l M E T E R Z E R O .
a.
Turn instrument on.
Allow at least a 20
minute warm-up period.
b. Turn voltmeter off, and allow 30 seconds for
all capacitors to discharge.
c.
Rotate mechanical zero-adjustment screw on
front panel clockwise until pointer reaches zero,
moving up scale.
d. If for
some
reason the pointer should over-
shoot zero, repeat step c until desired results
are obtained.
e. When pointer has been positioned at zero,
rotate zero-adjust screw slightly counterclock-
wise to free it. If meter pointer moves to the
left during this action, repeat steps c and e.
5-10. DC VOLTMETER OPERATION.
5-11. ACCURACY CHECK (DCV).
Model 738AR
a. Set the Model 410C FUNCTION SELECTOR
to the +DCV position; RANGE switch to. 015 V .
Connect Model 410C DCV and COM cables to the
Voltmeter Calibrator
Model 738BR) output
preferred instrument for these operations.
terminals.
Figure 5-1. Alternate Voltage Source
01556-2
5-1
Summary of Contents for 410C
Page 6: ...Figure 1 1 TM 11 6625 1614 15 M o d e l 4 1 0 C Section I Figure 1 1 1 0 ...
Page 8: ......
Page 9: ...Table 1 1 TM 11 6625 1614 15 Section I Table 1 1 Model 410C 01556 2 1 3 ...
Page 22: ...Figure 3 5 TM 11 6625 1614 15 Section III Figure 3 5 Model 410C 3 8 0 1 5 5 6 2 ...
Page 29: ...Figure 4 3 Model 410C TM 11 6625 1614 15 Section IV Figure 4 3 01556 2 4 3 ...
Page 30: ...Figure 4 4 TM 11 6625 1614 15 Section IV Figure 4 4 Model 410C 4 4 01556 2 ...
Page 31: ...Figure 4 5 TM 11 6625 1614 15 Section IV Figure 4 5 01556 2 4 5 ...
Page 32: ...Figure 4 6 TM 11 6625 1614 15 Section IV Figure 4 6 Model 410C 4 6 01556 2 ...
Page 33: ......
Page 44: ......
Page 47: ...Table 5 8 M o d e l 4 1 0 C TM 11 6625 1614 15 5 1 3 ...
Page 53: ...Figure 5 6 1 TM 11 6626 1614 15 Change 1 5 16 1 ...
Page 54: ......
Page 55: ...TM 11 6625 1614 15 Section V Figure 5 8 Figure 5 8 Power Supply Schematic 5 19 ...
Page 56: ...5 20 TM 11 6625 1614 15 Figure 5 9 Typical Amplifier Waveforms ...
Page 57: ......
Page 58: ......
Page 59: ......
Page 61: ......
Page 62: ...THE METRIC SYSTEM AND EQUIVALENTS ...
Page 63: ...PIN 016288 000 ...