(hp] MANUAL CHANGES
-hp- MODEL 400E/EL
AC VOLTMETER
Manual Part Number 00400-90020
Pigi 5·4. Tlbll 5·3.
Replace Table 5-3. Calibration
Tolerances. with Table MC1 of this Change Sheet.
Table Me 1.
1 Millivolt Ringe Only
Frequlncy
(Hz)
Voltlge
Meter (mV)
DC Output
Input
1%
of rllding)
(Volts)
10
1.00 mV
1.00
±
5%
1.00
±
0.05
0.5 mV
0.50
±
7.6%
0.50
±
0.038
0.3 mV
0.30
±
10.8%
0.30
±
0.033
40
1.00 mV
1.00
±
1%
1.00
±
0.02
0.5 mV
0.50
±
2%
0.50
±
0.015
0.3 mV
0.30
±
3.3%
0.30
±
0.013
100
1.00 mV
1.00
±
1%
1.00
±
0.005
or
0.5 mV
0.50
±
2%
0.50
±
0.0045
400
0.3 mV
0.30
±
3.3%
0.30
±
0.0043
CHANGE NO.5 Ipplies to III serial numbers.
Tlble5·1.
Required Test Equipment should have the follow
ing changes made:
a.
The AC Calibrator recommended model
should be a Fluke 5200A and 521 SA .
b.
The AC/DC Voltmeter recommended model
should be an -hp- Model 3466.
c.
The DC Standard recommended model
should be a Systron Donner Model
M
1 07 A
d .
The Voltmeter Calibrator should b e deleted
from the table.
Page 5·2. Figure 5·1.
Accuracy Test Setup, should be altered
to show the use of the Fluke 5200A ACk Calbrator rather
than the -hp-Model 745A .
Plge 5·4. Figure 5·2.
Accuracy and Frequency Response
Test Setup, should be altered to show the use of the
Systron Donner Model M 107 DC Standard rather than
the -hp- Model 740A.
Pege 5·11. Figure 5·2.
Delete paragraphs 5-45. 5-46, and
5-47.
Plge5·12.
Delete Figure 5-5. Alternate Calibration and Fre
quency Test Setup.
25 June 1984
1 Millivolt Rlngl Only
Frequlncy
(Hz)
Voltlge
Meter (mV)
DC Output
Input
1%
of reading)
(Volts)
100 k
1.00 mV
1.00
±
1%
1.00
±
0.005
0.5 mV
0.50
±
2%
0.50
±
0.0045
0.3 mV
0.30
±
3.3%
0.30
±
0.0043
500 k
1.00 mV
1.00
±
1 %
1.00
±
0.02
0.3 mV
0.30
±
3.3%
0.30
±
0.013
4 M
1.00 mV
1.00
±
5%
1.00
±
0.05
0.3 mV
0.30
±
10.8% 0.30
±
0.033
ERRATA
Plge &·3. Tlbll &·3 IRepllceable PlrtS).
Change the -hp- Part
Number of the A 1 Assembly from "00400-66521" to
"00400-66522" .
CHANGE NO. &. 400E: Applies to Seriel Numbers 1208A28944 Ind
Above.
4{JOEL: Applies to Serill Numbers 1208A29014 Ind
Above.
Section JVI, Tlble &·3 (Repllceable PlrtS).
Do the following
changes in the table:
A.f.
-hp- Plrt
Des.
IUlllber
Oescrlplion
A2C47
0180-0100
Add 4.7/AF 35V
Capacitor
A2Q17
1853-0010
Add SM4713
PNP
Transistor
A2Q18
1855-0093
Add F17 48
N
Channel
JFET
A2R76
0683-1 045
Add lOOK
ohms
5% 1/ 4W
Resistor
A2R77·
0757-027 4
Add 1 . 1 2K
ohms
1 % 1/8W
Resistor
A2R77"
0698-3700
Add 7 15
ohms
1 % 1/8W
Resistor
A2R77·
0698-0410
Add 301
ohms
1 % 1/8W
Resistor
A2R78
0698-3458
Add 348K
ohms
1 % 1/8W
Resistor
Section VII. Figurl 7-1 (400 EIEL Schlmetic Dilgrlm Ind locltion
of Componlnts).
Do the following changes in the figure.
Change the Meter Bridge to the bridge shown in Figure
C-1.
Change the component locator of the A2 Assembly (-hp
Part Number 00400-66521) to the one shown in Figure
C-2.