![HP 400E Operating And Service Manual Download Page 31](http://html.mh-extra.com/html/hp/400e/400e_operating-and-service-manual_161442031.webp)
Models 400E/4DOEL
e. Place the oscillator and the 400E/EL on the
I m V r ange. Increase the oscillator's
amplitude for full scale deflection on the
400E/EL meter. The ac digital voltmeter
should read I05mV ±IO%.
5-31.
C ONCLUS ION
O F P E RFO RMANC E
CHECKS.
NOTE
When the 400E/EL has passed these
checks it has met its specifications
listed in Table
1-1.
5-32.
A LIGNMENT AND CALIB RAT ION
P ROCEDU RE.
NOTE
The location of adjustments on A2
depends on the applicable board
revision (see Pages 7-2 and 7-3).
5-33. The calibration adjustments are "cover off'
procedures to adjust the 400E/EL to its performance
specifications. I f the instrument cannot be properly
adjusted, refer to the Troubleshooting Procedures
(Paragraph 5-48). Figure 5-4 shows the location of all
the internal adjustments.
Section V
5-34.
COVE R REMOVAl.
5-35. To remove the top or bottom covers, remove
the Phillips screws holding the cover, slide the cover
about
1
inch to the rear, and lift it off. To replace the
cover, reverse the removal procedure. If it is necessary
to remove a side cover, remove the four Phillips
screws and lift it off.
Connect a dc voltmeter to TPI. The
vol t age should be +26 ±2V.
Connect a dc vol tmeter to TP2. The
voltage should read -26 ±2V. If
these voltages are not correct refer
to Troubleshooting Paragraph 5-48.
5-36.
BIAS ADJUST .
5-37. Connect a dc voltmeter to TP3 and adjust
A2RI7 for -6.0 ±0.25V dc. Connect a dc voltmeter to
TP4 and adjust A2R31 for
+
IO.O ±I V dc.
5-38.
AC OUTPUT ZE RO.
5-39. Connect a dc voltmeter to TP5 and adjust
A2R59 for 0.0±0.050Vdc.
5-40.
CA II B RA TI
0
N.
I f a 400E/EL Option 02 is to be
A2R38
400
Hz
ADJ
A2C28
A2R31
S2C2
A2R59
OV
ADJ
A2R44
10
MHz
ADJ
BIAS ADJ
,------...
,
I
R2
1 mV
400
Hz
A DJ
DC OUTPUT ADJ
(ADJ FROM SIDE)
3
mV
10
MHz
ADJ
(ADJ FROM SIDE)
A1C2
3
V 100 kHz
ADJ
(ADJ FROM SIDE)
Figure 54. Location of Internal Adjustments