With   the   function   [Change   of   bath],   the   chosen   bath   is   placed   facing   the   user   to   allow   the   changing.   For   
safety   reasons,   the   door   must   be   kept   closed.   If   not,   the   machine   will   ask   its   closure.   In   this   position,   a   
buzzer   sounds,   the   door   can   be   opened   and   the   bath   changed.         
   
After   the   changing,   the   cursor   automatically   jumps   on   the   function   [Reset   of   the   counter],   whereby   both   
counters   can   be   reset   to   zero.   In   the   window   "Bath   cycles",   one   can   enter,   under   point   "Maximum",   the   
maximum   number   of   cleaning   cycles   that   the   bath   can   use.   The   using   bar   generates   itself   on   the   main   
screen.   The   counter   takes   also   in   account   the   previous   uses.   
   
In   the   window   "Age   in   days",   one   can   set   the   maximum   duration   of   use   in   days.   Similar   to   the   cleaning   
cycles,   the   consumption   bar   is   generated   since   the   last   bath   changing.   
   
   

CAUTION   

:   If   for   both   "Cleaning   cycles"   and   "Age   in   days"   a   maximum   value   is   defined,   the   consumption   bar   

displays   the   value   that   comes   first   to   the   limit   or   the   one   that   is   closest.   

   

   
   
If   during   cleaning,   the   maximum   use   of   a   bath   is   reached,   an   indication   appears   after   cleaning.   After   
confirmation,   the   menu   goes   automatically   to   summary   screen   of   the   baths.   Here   it   is   possible   to   identify   
the   bath   which   limit   has   been   reached   and   change   it.   
   
In   the   field   "Name   of   the   bath",   it   is   possible   to   name   the   bath   with   its   cleaning   fluid.   The   shortcut   "B1"   
refers   to   the   position   1   of   the   bath   in   the   machine.   If   no   name   is   entered,   "The   name   of   the   bath"   field   will   
be   blank   and   the   default   name   appears   or   only   the   position   if   the   name   has   been   deleted.   

Using   the   arrow   at   the   top   right,   one   can   return   to   the   overview   of   the   baths,   and   according   to   the   same   
principle,   all   positions   of   baths   can   be   started,   adjusted   and   changed.   
   
After   leaving   the   bath   changing   menu,   the   machine   performs   a   homing   of   the   axis.   

   

   

   

   

   

   

   

   

   

   

   

Summary of Contents for VIBRATO MSA 19.709

Page 1: ......

Page 2: ...ing this option the VB movement can be selected and customized in the settings During this oscillating motion vibrating movement it is possible to constrain the rotation angle to 36 degrees MSA19 710...

Page 3: ...age 5 5 Safety instructions Page 6 6 Technical data Page 7 7 Starting Pages 8 9 8 Operation Page 10 9 Equipment Pages 11 13 10 Creating cleaning programs Pages 14 18 11 Performing cleaning programs Pa...

Page 4: ...cm for organic solvents The amplitude of the depression to be supplied to reach the cavitation threshold depends on several parameters the higher the viscosity of the media therefore the internal coh...

Page 5: ...economical an alternative to solvents biodegradable detergents reduced cleaning time accurate effective and easy 4 EC declaration of conformity This device complies with European standards CE accordi...

Page 6: ...e of the unit by any other person is not possible Make sure that the machine operates only in a well ventilated room or connected to an external exhaust Take care when loading the machine not to damag...

Page 7: ...packing 102 kg Dimensions 480 x 480 x 760 mm Country of origin Germany Specific technical shipping information Extracted air Internal 80 mm Suction force 26 qm h free movement Standard content of del...

Page 8: ...ered in winter at very low temperatures the formation of condensation can damage the electronics In this case and before use leave the device at room temperature for 24 hours to avoid any risk of dama...

Page 9: ...achine to the power supply 2 Switch on the machine by pressing the On Off switch to 1 The unit performs a first self test and starts the initialization This means that all axes will go to the referenc...

Page 10: ...r looks like two brackets around the value to be modified Value To change a value press briefly on the ENC which changes the brackets around the value Value Turning the ENC can now change the value it...

Page 11: ...can rotate freely The wave breakers placed in the tanks prevent rotation of the fluid 2 Fill the tanks up to the mark with the desired cleaning fluid Beware of the instructions for use concerned fluid...

Page 12: ...he position of the individual tank on the turntable and starts with the ultrasonic bath B1 counter clockwise This allows the identification of the tank and its position on the turntable which cannot b...

Page 13: ...ing CAUTION If for both Cleaning cycles and Age in days a maximum value is defined the consumption bar displays the value that comes first to the limit or the one that is closest If during cleaning th...

Page 14: ...very position All the steps can be programmed in any order and are not related to the sequence of the baths in the machine Similarly individual tanks can be repeatedly used with different cleaning mov...

Page 15: ...determined on the basis of different cleaning times in each bath spinning and drying times as well as the duration of the movements As soon as a cleaning program has been completely executed the machi...

Page 16: ...tep select the field New step and confirm with the ENC Left you see the statement on the menu It is also possible to insert a new step at any point of an existing program with the START STOP button Af...

Page 17: ...is will be explained later in this manual The following movements are available LD SD LW SW LO SO LX SX VB optional 03 00 Setting the duration of the bath This means how long the basket in movement li...

Page 18: ...he PTC heating element Please make a test run without any pieces after each temperature setting When all the cleaning cycle steps are performed one can access to the summary of the program by moving t...

Page 19: ...sically the machine starts with the last executed or changed cleaning program To select a different program turn the ENC until the name of the program is highlighted or black and confirm the selection...

Page 20: ...s ALWAYS by pressing the Start Stop button to prevent accidental starts during the manipulation with the ENC Interrupt the cleaning program If the Start Stop button is pressed during an active cleanin...

Page 21: ...om the fluid and the spinning cycle begins if programmed before the next step starts If the cleaning basket had no contact with the fluid the spinning movement is skipped and the step begins immediate...

Page 22: ...basket for each bath and the degrees of rotation of the simple movements if enabled To try this option press the Start Stop button Here all settings can be made For example the rotational speed of th...

Page 23: ...ns left then 0 7 n turns right fast acceleration vibrasonic movement Now one can globally set the rotation speed for every single movement For oscillating or changing movements the optional software a...

Page 24: ...s the currently unlocked functions Furthermore the other functions can be unlocked To unlock an optional function please contact us to get the code The machine generates a new request code REQ at ever...

Page 25: ...as to be controlled and if necessary applied some grease To do this we recommend you the spindle grease reference L1 ZBH003 applied with a bristle brush through the slots in the front panel Check air...

Reviews: