44 1538 101-B
Manual flow meter FMOGne
9
between “Litre”, “US-Gal” and “UK-Gal”. Renewed pressing of the “Mode” key and the
set measurement unit is taken on. Notice: Any change to the measurement unit causes
the volume indicator and totaliser to be reset.
5.4
Setting the calibrating factor
After pressing the “Mode” key, the calibrating factor set appears in the display.
Pressing the “Total” key results in the calibrating factor being counted up in 0.010
steps and pressing the “Reset” key in it being counted down accordingly.
Press the “Mode” key and the set calibrating factor is taken on.
5.5
Resetting all settings (initialisation)
The effect of pressing and holding the “Total” and “Reset” keys at the same time when
setting the measurement unit or calibrating factor is to delete all the meter values
and carry out an initialisation.
Following values are initialised:
Measurement unit:
Litre
Calibrating factor:
1.000
Delivery quantity:
0.00 litres
Totaliser:
0 litres
5.6
Ending the programming mode
To end the programming mode, the “Mode” key is again to be pressed after setting
the calibrating factor. The meter then reverts to the basic state. When no key is
pressed for over 5 minutes in the programming mode, the meter automatically
returns to the basic state.