TDS set up
28
●
P2 TDS setup
Using P2 TDS setup function of the meter, you can:
• Select TDS curve
• Select TDS unit
To set the TDS functions using LAQUA EC200 handheld meter, follow the procedure detailed
below:
Prerequisites
Switch on the EC meter.
• Default TDS curve is
linear
. You can change the TDS curve to 442 or EN27888 or NaCl.
• For linear curve, default multiplier factor is
FACT 0.50.
You can set a multiplier factor in
between 0.40 to 1.00.
• Default TDS unit is set as
mg/L
(g/L). You can change the unit to ppm (ppt).
To return to the measurement mode, press the
key.