Conductivity calibration
12
Calibration
Auto calibration
1. After placing the conductivity electrode in the standard solution, press the
key.
2. Meter displays “Auto cal” as per set calibration method and starts measuring various
calibration values with a blinking
on screen.
3. Wait for the
to stabilize (stable calibration reading).
4. Press the
ENT
key to confirm and save calibration data.
5. Meter displays
DONE
indicating end of the conductivity calibration procedure.
6. Repeat for other calibration points as required.
7. You can calibrate at one point for each range
If you want to know previous calibrated values, press the
key when you are in the
CAL
mode.The display scrolls through the calibrated values.
Note