Temperature Calibration
16
Calibration
1. After placing the electrode in the solution, press the
key.
2. Press the
key to switch to temperature calibration mode. Meter displays measured
temperature value.
3. Use the
keys to adjust the temperature to the required value.
4. Press the
ENT
key to save calibration data.
5. Meter displays
DONE
indicating end of the temperature calibration procedure.
key
key
key
key