pH Calibration
11
■ Calibration
This section describes the calibration procedures using the LAQUA PH1500 bench meter and
electrodes.
● pH Calibration
Calibration is necessary for accurate pH measurement. To perform pH calibration, follow the
procedure detailed below:
Prerequisites
• Clean the pH electrode with DI (deionized) water and wipe it with tissue paper.
• Switch on the pH meter and plug in the pH electrode.
• Prepare buffer solution required for calibration.
• Keep the meter in pH measurement mode.
• Dip the pH electrode at least 3 cm in the buffer solution.
• Perform two-point calibration using:
pH 7 and 4 for acidic sample.
pH 7 and 10 for alkaline sample.
• Perform three-point calibration using pH7, 4 and 10 if you are unsure of the expected
sample pH value. It is recommended to calibrate with pH7 first.
• Default buffer setup is
BUFF USA
. If you want to change to
BUFF NIST
or
BUFF DIN
,
refer to “P 1.1 Buffer selection” on page 20.
• To abort an ongoing calibration process at any point of time, press the
key.
• It is recommended to clear the previous calibration data before performing calibration. For
erasing the calibration data, refer to ““P 1.3 Erase calibration data” on page 22.
Note
Tip
Summary of Contents for LAQUA-PH1500
Page 1: ...Instruction Manual Benchtop pH ORP Meter LAQUA PH1500...
Page 2: ......
Page 4: ...MEMO II...
Page 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Page 23: ...Product Overview 6...
Page 49: ...General Setup 32...
Page 53: ...Maintenance and Storage 36...
Page 63: ...Meter Specifications 46...