7
●Ion calibration start command
Auxiliary unit of calibration value unit to be set.
0:-
1:μ
2:m
Follows the unit setting for mol/L or g/L.
Calibration value to be set
Channel (1 or 2)
Name
Header
Starts the ion calibration when waiting for ion calibration.
ER, 2 is returned when the temperature value is Or (Ur).
Calibration is possible when the measurement value is Or (Ur).
Setting range of calibration value
Unit: mg/L, μg/L, mmol/L, μmol/L
0.00 to 9.99
(Possible at [SP][SP]X or [SP]X.X.)
10.0 to 99.9
(Possible even at [SP]XX.)
[SP]100 to [SP]999
※ Automatically returns to the measurement mode after the calibration is finished.
※ Automatically returns to the measurement mode even if a calibration error is generated.
※ A calibration error can be checked by the alarm inquiry (R, AL).
※ Calibration errors are released by the alarm clear (R, AR).
C,
CI
,
x
,
yyyy
, z
[CR][LF]