6
●Calibration mode command
0: Calibration mode end, 1: Calibration mode start
Channel (1 or 2)
Name
Header
Waits for the calibration at specified mode.
※ To return to the measurement mode without calibrating after entering the calibration mode,
enter "0: Calibration mode end".
●pH calibration start command
Calibration value to be set (when custom calibration is set)
Channel (1 or 2)
Name
Header
Starts the calibration when waiting for pH calibration.
Even when the setting is other than the custom calibration,
it is necessary to enter the calibration value (It will not be used.).
ER, 2 is returned when the measurement value or temperature value is Or (Ur).
Setting range of calibration value
Specified by three decimal places.
0.000 to 14.000
※ Automatically returns to the measurement mode after the calibration is finished.
※ Automatically returns to the measurement mode even if a calibration error is generated.
※ A calibration error can be checked by the alarm inquiry (R, AL).
※ Calibration errors are released by the alarm clear (R, AR).
C, CM,
x
,
y [CR][LF]
C, CP,
x
,
xxxxxx [CR][LF]