20
Response from pH meter
Second 2 digits 00 to 59
Minute 2 digits 00 to 59
Hour 2 digits 00 to 23
Day 2 digits 01 to 31
Month 2 digits 01 to 12
Year 4 digits A.D.
Ion type 1 digit
For ion measurement
0: -2 1: -1 2: +1 3: +2
For other measurement parameter
[SP]
Measurement state 1 digit
0: Instantaneous value 1: Hold state
2: Potential follow-up measurement
Measurement/calibration type 1 digit
0: Measurement 1: Calibration
Channel 1 digit
1: CH1 2: CH2
Measurement mode 1 digit
1: pH 2: Absolute mV 3: Relative mV 5: Ion 10: Conductivity
12: Resistivity 11: Salinity 13: TDS
Sample ID 4 digits
※Enter spaces.
[SP][SP][SP][SP]
Header
Error state
2: Upper limit alarm
1: Lower limit alarm
0: No alarm
Potential 7 digits
(Leading zero suppression)
Temperature (6 digits including decimal point) -30.0 to 130.0
Right-justified
When the temperature value is Or (Ur), the character string
Or or Ur is displayed instead of an upper or lower limit value.
Temperature setting 1 digit
0: ATC 1: MTC
Unit 1 digit
(pH)
0: Non
(ORP)
0: mV
(Ion)
0: μg/L, 1: mg/L, 2: g/L, 3: mmol/L, 4: mol/L
(Conductivity)
0: S/m, 1: S/cm, 2: mS/cm-Fix
(Resistivity)
0: Ohm/m, 1: Ohm/cm
(Salinity)
0: PPT, 1: %
(TDS)
0: g/L
Auxiliary unit 1 digit
0: Non 1: μ 2: m 3: k 4: M
Measurement value max. 7 digits
RMD,
xxxx
,
xx
,
x
,
x
,
x
,
x
,
xxxx
,
xx
,
xx
,
xx
,
xx
,
xx, (⇒)
(⇒) Xxxxxxx
,
x
,
x
,
x
,
xxxxx
,
xxxxx
,
x [CR][LF]