Chapter 3 Basic Operation
3.2 Settings
34
HORIBA
●
Temperature Calibration
You can perform calibration of the temperature sensor of
an electrode.
1.
Press the SET key on the measurement screen,
and press the
▲
or
▼
key to enter the maintenance
screen.
2.
Press the ENTER key to enter the maintenance
item selecting screen.
3.
Press the
▲
or
▼
key to enter the temperature cali-
bration screen.
4.
Press the ENTER key to enter the temperature dis-
play value change screen.