Chapter 4 Maintenance and Troubleshooting
4.2 Troubleshooting
54
HORIBA
4.2
Troubleshooting
This instrument is equipped with a simple error-message
function to notify the operator that an operation error or
problem with the equipment has occurred. Errors or
other problems that occur in the measurement mode are
announced by an "ERROR No." appearing in the upper
lefthand corner of the screen.
4.2.1 Error Message Chart
ERROR
No.
Message
Definition
0001
Memory ERROR
Data cannot be read from or written to
the internal memory.
0003
Unstable reading
ERROR
The electric potential does not stabilize
within three minutes.
0009
Printer
communication
ERROR
There is a problem with the printer com-
munication.
0010
Data memory over
The number of the memory data
exceeds the limit of the memory.
0011
Cell constant out of
range
The c
COND:
ell constant is out of the
automatic calculation range.
000E
Key operation
ERROR
Key operation is invalid.