Chapter 2 Taking Measurements
2.5 Conductivity measurement
20
●
Measuring conductivity
1.
Immerse the electrode in the sample.
Conductivity is greatly affected by temperature.
To measure with increased accuracy, use a temperature bath to keep the
solutions at a constant temperature.
2.
Press the MEAS key while the Instantaneous Value Measurement
screen is displayed.
The measured value will be displayed, and
“
HOLD
”
will blink until
the reading stabilizes. When the measured value stabilizes,
“
HOLD
”
will stop blinking and the measured value will remain
displayed, and measurement will be completed.
Refer to the “ Criteria for judging stability” page 14 for the criteria for
judging the stability of the readout.
When the COND meter is in the Instantaneous Value Measurement mode or
the measurement value is on HOLD in the Auto Hold Measurement mode, you
can store the measurement data to memory by pressing the
▲
key. See