Chapter 2 Taking Measurements
2.5 Conductivity measurement
18
●
CELL SET mode (Setting cell constant)
Set the cell constant the first time an electrode is connected to the main
unit of the meter.
1.
To enter the CELL SET mode, press the CAL key while in the
Measurement mode.
2.
Change the digit number using the ENTER key.
3.
Press the
▲
and
▼
keys to set the cell constant written on the
electrode label.
Setting range: 0.700 – 1.300
To change the coefficient, use the following procedure.
When the SI unit system (m
-1
) is set:
When the former unit system (cm
-1
) is set:
Temperature coefficient
The default value of the temperature coefficient is set at 2.00%/ºC.
To change this setting, refer to
Unit Setting
The default value of unit is S/m (SI unit system).
To change this setting to the former unit system S/cm, refer to
3.3.3 Conductivity unit setting
P.29.
Digit