Chapter 3 Functions
3.1 Data memory function
DS-51
23
Chapter 3
Functions
This chapter describes the various functions of the COND meter.
3.1
Data memory function
The measured data can be stored manually.
●
Data memory
In all measurement modes, you can store data when the instantaneous
value is measured or the measured value is held (HOLD status) during
the Auto HOLD measurement by pressing the
▲
key.
The measurement reading is stored along with the temperature, data,
HOLD value/instantaneous value, ATC/MTC, calibration point (pH), and
sample ID at the time the measurement was taken.
After the data number is displayed, the screen returns to the
Instantaneous Value Measurement mode. Up to 300 items of data can
be stored in the memory. If the number of data items exceeds the
maximum limit, ERR 10 is displayed and no more data can be stored.
Data cannot be stored unless the value has stabilized or in the CAL mode.
When the data is stored, an ID number for that specific measurement
can be registered (see page 30).