horiba 9625-10D Instruction Manual Download Page 8

4

Maintenance

Note

Do not use organic solvents (such as acetone or
tetrahydrofuran) or acid (above 1 mol/L) or alkaline (above 0.1
mol/L) cleansing solutions. Using such chemicals may 
damage the body of the electrode or cause a decline in
performance.

To maintain electrode performance, replace the internal
solution approximately every one or two months. 
(Refer to " Preparation " (page 2).)

Dirt on the responsive glass membrane or liquid junction
by the test solution may cause a decline in electrode
response and sensitivity or measurement error. If the
electrode is very dirty and cannot be washed clean
using pure water, perform cleansing.

Cleansing of electrode

1. Open the internal solution filler port.
2. Immerse the responsive glass membrane and liquid

junction in cleansing agent for approximately one hour
until contamination is removed. The responsive glass
membrane can also be cleaned by wiping it with cotton
gauze containing cleansing agent.

3. Rinse the electrode with pure water completely.

4. Replace the internal solution.

(Refer to " Preparation " (page 2).)

Note

Be sure to perform calibration prior to measurement after
cleansing the electrode.

Tip

Applying pressure on the internal solution and exuding it from
liquid junction have an effect of stabilizing pH response.

Type of dirt

Cleansing agent

General

Diluted neutral cleansing agent

Oil

Alcohol, or diluted neutral cleansing agent

Inorganic substance

Electrode cleaner (model 220) or Hydrochloric 
acid approximately 1 mol/L

Protein

Cleansing agent including protein-cutting 
enzyme (model 250)

Liquid junction

Electrode cleaner (model 220)

Altered of glass, 
Slow-response

Electrode cleaner (model 220)

Storage

1. Wash the electrode well with pure water to remove

sample completely and wipe it with filter paper.

2. Close the internal solution filler port.

3. Wash the inside of the protective cap with pure water,

then add enough pure water to soak the sponge.

4. Attach the protective cap.

Note

Do not dry the responsive glass membrane and liquid
junction. If they are dry, replace the internal solution and
immerse them in pure water for 24 hours or longer. After
these procedures, perform calibration. If calibration cannot
be performed correctly, replace the electrode.

Avoid storing the electrode in hot and humid locations.
Store the electrode indoors, out of direct sunlight.

For accurate measurement

For accurate measurement, refer to items below.

Stir the sample with stirrer in measurement (calibration).

Perform temperature compensation manually or
automatically. 
(Refer to the instruction manual of pH meter.)

Temperature of standard solution and sample should be
the same.

Rinse the electrode with the sample before
measurement.

Immerse the responsive glass membrane and liquid
junction in the sample completely. Make sure that the
surface of the internal solution inside the electrode is
higher than the surface of the sample solution.

Perform calibration before measurement.
If measurement is performed everyday, perform
calibration once or more a day.

Protective cap

Reviews: