horiba 9625-10D Instruction Manual Download Page 7

3

Measurement (calibration)

1. Open the internal solution filler port.

2. If necessary, fill the electrode with new internal solution

until the solution level gets close to the internal solution
filler port.

3. Wash the electrode with pure water and wipe it with filter

paper, prior to every immersion into standard/sample
solution.

4. When there are air bubbles inside the responsive glass

membrane, stir it lightly to remove bubbles.

5. Immerse the responsive glass membrane and liquid

junction in the sample completely. 
Make sure that the surface of the internal solution inside
the electrode is higher than the surface of the sample
solution.

6. Stir the electrode lightly to remove bubbles.

Note

Be sure to keep the internal solution filler port open during
measurement. If it is clogged with the solution, accurate
measurement cannot be performed. Wipe the solution with
cloth.

For calibration/measurement on pH meter, refer to the
instruction manual of pH meter.

In submergence measurement

Perform measurement with the internal solution filler
port closed.

Pour the internal solution as much as possible, and
close the internal solution filler port not to leave the
bubbles inside the electrode.

Avoid continuous use for more than 30 minutes.

Do not swing around the sensor, and do not pull the
connection with the cable. In addition, do not hold the
cable the electrode and pH meter hanging. It may cause
breaking of wire.