Before use, immerse the sensor into 3.33 mol/L KCl solution (300) for 30 minutes or longer.

6.

Storage

7.

Maintenance

Cleansing the sensing part

Replacing the sensor

If any of the following symptoms is found after responsive-part cleaning, replace the sensor in the procedures below.

Sensitivity or response speeds do not recover.

The electrode can not be adjusted in the way mentioned in "4. Calibration" (page 10 ). 

The electrode can not be calibrated with standard solutions. 

Replacing batteries

Note

Replace two batteries with new ones (CR2032) at the same time.  

Do not touch the sensor terminal. It may be damaged by static electricity.

Make sure to remove the sensor prior to the converter cover. Removing the cover with the sensor attached causes
malfunction.  

1

Press the power switch to turn OFF the power. The Meas. LED turns off. 

2

Wash the electrode well with pure water to remove sample completely and wipe it 
with filter paper,  and store it in dry conditions.

Note

Avoid storing the electrode in a hot place or a highly humid place. 
Keep the electrode dry and out of direct sunlight. 

Dirt on the sensing part may affect the electrode response, sensitivity, or a mea-
surement precision. If the electrode is very dirty and cannot be washed clean using 
pure water, clean it as follows.

1

Clean the sensing part by wiping it with cotton gauze containing cleans-
ing agent.

2

Rinse the electrode with pure water completely. 

Note

Be sure to calibrate the electrode after cleansing the sensing part.

    

Type of dirt

Cleansing agent

Note

Do not use organic solvent (e.g. acetone, THF), strong acid or 
alkali. Using these chemicals may damage the electrode.

General

Diluted neutral cleansing agent

Oily

Diluted neutral cleansing agent, or alcohol

Inorganic (AgCl, etc.)

Cleansing agent (220)

Turning off

 Power switch

1

Turn OFF the power.

2

Turn the nut to the left to remove it. And pull out the 
sensor from the converter.

3

Attach a new sensor referring to "

 Assembling sen-

sor and converter" (page 9).

 

Note

Do not touch the sensor terminal.  It may be damaged by static electricity.

When battery voltage is low, the Low Batt. LED blinks red. 
(The LED is off during battery shutoff.) In this case, replace 
batteries as follows:

1

Turn OFF the power.

2

Turn the nut to the left to remove it. And pull out the 
sensor from the converter.

 

3

Replace batteries with new ones referring to "

 Assembling sensor and converter" (page 9).

If used batteries are set, remove it by using a ball-point pen or the like.

4

Attach the sensor referring to "

 Assembling sensor and converter" (page 9)

 Turning off

Power switch

Nut

Turning off

Power switch

Nut

Summary of Contents for 0030-10D

Page 1: ...charge should any mal function occur during the course of normal use if returned within the warranty period 12 months from date of purchase Return the con verter with proof of purchase date This warra...

Page 2: ...INSTRUCTION MANUAL 2 1 ISFET 70 mm 0 01 mol L ABS 1 3 33 mol L KCl 300 30 0 60 C 0 40 C ISFET 6 25 C 200 OFF Low Batt LED Meas LED 500 15 70 mm...

Page 3: ...0131 pH 0 14 0 60 C 0 40 C 0 40 C 190 mm ISFET CMC KCl 5 ABS 16 mm pH D 10 D 20 D 50 D 70 F 10 F 20 F 50 F 70 DC 6 V CR2032 2 OFF Meas LED ON Low Batt LED 1 m 0131 1 1 CR2032 2 3 33 mol L KCl 300 1 1...

Page 4: ...INSTRUCTION MANUAL 4 3 ISFET pH Note ISFET pH pH 1 pH 2 3 pH 4 pH 1 2 CR2032 2 3 a A a A 1 O 2 3 O 4 3 33 mol L KCl 300 30 O...

Page 5: ...TION MANUAL 5 4 Note ISFET 5 Note 3 33 mol L KCl 300 30 1 3 33 mol L KCl 300 10 2 ON Meas LED Note ON 1 OFF 3 pH7 4 pH Note pH pH CLR 5 pH pH 1 2 ON Meas LED Note ON 1 OFF 3 4 pH mV 0 10 mV pH pH 7 pH...

Page 6: ...INSTRUCTION MANUAL 6 6 7 4 5 Note 2 CR2032 1 OFF Meas LED 2 Note 1 2 Note Note 220 1 OFF 2 3 4 Note Low Batt LED 1 OFF 2 3 4 4 4...

Page 7: ...can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a resi...

Page 8: ...sensor cannot be repaired Before replacing the sensor or batteries turn off the power and wipe water off the electrode by patting the electrode dry with filter paper Make sure to detach the sensor be...

Page 9: ...he nut is not tightened enough water may enter the inside of the sensor and the sensor may be damaged Before measurement or calibration prepare the ISFET pH electrode as follows Connecting to pH meter...

Page 10: ...pH7 standard solution with the sensing part and liquid junction completely immersed If necessary swing the sensor gently in the solution to remove bubbles near the sensing part and liquid junc tion 4...

Page 11: ...If the electrode is very dirty and cannot be washed clean using pure water clean it as follows 1 Clean the sensing part by wiping it with cotton gauze containing cleans ing agent 2 Rinse the electrod...

Page 12: ...i cho Kisshoin Minami ku Kyoto 601 8306 Japan 0120 37 6045 9 00 12 00 13 00 17 00 For any questions regarding this product please contact your local agency or inquire from the following website http g...

Reviews: