Preparation
12
■
Preparation
●
Confirmation before starting measurement
・
Have you prepared the appropriate electrode for the measurement parameter?
⇒
If not, purchase the appropriate electrode.
・
Is the prepared electrode in good condition?
⇒
If the responsive part is stained or damaged, it may not be possible to obtain
accurate values.
・
Have you prepared the appropriate standard solution for the measurement parameter?
⇒
If not, prepare the standard solution by yourself or purchase it.
・
Are there any items that should not be wet or stained around the instrument?
⇒
Depending on the operation during measurement, items around the instrument could
get wet or stained. Secure sufficient space around the instrument and perform
measurement while always paying attention to safety.
・
Are there any devices that can be a source of noise?
⇒
Measured values could be affected. Do not use the instrument near such devices.
Always ground devices operated by AC power.