Instrument default settings
84
■
Instrument default settings
Measurement
parameter
Item
Description
Dissolved
oxygen
Measuring principle
Membrane galvanic cell
Measuring range
(Display range)
0.00 mg/L to 20.00 mg/L
Resolution
0.01 mg/L
Repeatability
±0.1 mg/L ±1 digit
Oxygen
concentration
Measuring principle
Conversion from dissolved oxygen value
Measuring range
(Display range)
0.0% to 50.0%
Resolution
0.1%
Saturated
oxygen
concentration
Measuring principle
Conversion from dissolved oxygen value
Measuring range
(Display range)
0.0% to 200.0%
Resolution
0.1%
Measurement
parameter
Item
Selection item/Setting
range
Default values
Common
Temperature setting
Automatic (ATC)/
Manual (MTC)
Automatic (ATC)
Temperature input value
0.0°C to 100.0°C
25.0°C
Auto power off
ON/OFF
OFF
Auto power off time
1 min to 30 min
30 min
Sample ID
000 to 2999
000
Auto data memory
ON/OFF
OFF
Auto data memory time
2 seconds to 24 hours
2 seconds
pH
Standard solution
NIST/USA/CUST
NIST
Calibration interval
alarm
ON/OFF
OFF
Calibration interval
1 day to 400 days
1 day
Dissolved
oxygen
Salinity correction
Sea concentration (SEA)/
numeric setting
numeric setting
Salinity correction value
0.0 PPT to 40.0 PPT
0.0 PPT
Air pressure value
100 hPa to 1999 hPa
1013 hPa
Industrial standard
ISO/Pre
ISO