© 2014 Hopkins Medical Products
0
CAL
Calibration Procedure
2.
To enter the Calibration Mode,
press and hold both the
"LB/KG"
button and the
"ON/TARE"
button
simultaneously until the LCD
screen displays "
CAL
".
1.
Place the scale on a level and
very stable surface. It is important
for the scale to be stable and level
in order to preform an accurate
calibration.
Leave the scale turned off.
3.
The scale will begin to display
a series of calibration numbers.
These numbers will begin to
stabilize and will begin to fluctuate
by only a few digits. Press
the
"ON/TARE"
button when
the number is only fluctuating
minimally.
4.
The LCD screen will briefly
flash "
C10
" and then display "
0
".
5.
Place a 10kg certified weight
on the scale and wait until the
calibration numbers stabilize
again.
6.
Press the
"ON/TARE"
button
again once the numbers have
stabilized as before and are only
fluctuating minimally.
9
688329 Max=44lbs/20kg
d=0.01lbs/5g
688329 Max=44lbs/20kg
d=0.01lbs/5g
10 KG
*Calibration numbers shown in this manual are for illustrative purposes only.