XNX Universal Transmitter
Section 3 - Calibration
115
Note:
Calibrated mA output is required for proper operation of internal diagnostics.
NOTE
!
An F165 fault will be reported if the 4-20 mA calibration fails.
Align Excel (Searchline Excel)
Figure 221. Align Excel Menu
For detailed information on aligning the Searchline Excel, see
the
Searchline Excel Technical Manua
l (Honeywell part number
2104M0506).
Alignment Bargraph
Signal Strength
1.00 Signal Strength Target
Reference Value
Figure 222. Align Excel Screen
Align the unit using the information found in the Searchline Excel
manual. As the alignment is performed, the transmitter display will
indicate the signal strength in the form of a bar graph. Align the
Excel until the signal strength bar graph reaches or exceeds 1.00 as
shown on the display.
Soft Reset
(Searchline Excel and Searchpoint Optima Plus only)
Figure 223. Soft Reset Menu
For transmitters connected to a Searchline Excel or Searchpoint
Optima Plus sensor, the Soft Reset sends these infrared devices a
signal to restart the sensor.
Figure 224. Soft Reset Sensor Screen
Summary of Contents for XNX-AM Series
Page 2: ...This page is intentionally left blank ...
Page 7: ...XNX Universal Transmitter Safety and Information 5 Safety and Information ...
Page 12: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 10 Introduction ...
Page 27: ...XNX Universal Transmitter Introduction 25 Menu Navigation ...
Page 102: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 100 Calibration ...
Page 118: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 116 Maintenance ...
Page 122: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 120 Warnings and Faults ...
Page 139: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 137 Specifications ...
Page 174: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 172 Control Drawings ...
Page 177: ...XNX Universal Transmitter Control Drawings 175 ...
Page 179: ...XNX Universal Transmitter Control Drawings 177 Remote Sensor Mount ...
Page 180: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 178 HART Protocol ...
Page 193: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 191 Modbus Protocol ...
Page 200: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 198 Warranty ...
Page 202: ...XNX Universal Transmitter XNX Universal Transmitter Technical Manual 200 Index ...