background image

 

15 

3. 

Calibration 

The ToxiPro features fully automated fresh air/zero and 
span calibration functions.  The MODE button is used to 
initiate the automatic calibration sequence.  Calibration 
adjustments are made automatically by the instrument. 

3.1 

Verification of accuracy 

3.1.1  Verifying accuracy and 

response: ToxiPro O

To verify the accuracy of the ToxiPro O

2

, take the 

ToxiPro O

2

 to an area where the atmosphere is known to 

be fresh and check the readings.  If the readings differ 
from those expected in fresh air (oxygen monitors should 
read 20.9% in fresh air), then a fresh air/zero calibration 
adjustment must be made as discussed below in 
section 3.3.  If fresh air is not available, see section 3.7 
below for instructions for calibrating the ToxiPro O

2

 in 

contaminated air. 

Honeywell Analytics also recommends that the response 
of the oxygen sensor be regularly verified by any of 
these methods: 

 

Expose the O

2

 sensor to a known concentration 

of gas containing less than 19.0% oxygen.  If the 
descending oxygen alarm is set to 19.5% the 
instrument should go into alarm a few seconds 
after the gas reaches the sensor face. 

 

Process the ToxiPro O

in an IQ Express Dock 

that has a cylinder of calibration gas containing 
18.0% oxygen (or less) connected to the gas port 
and proceed with the oxygen bump test. 

 

Breath test:  Hold your breath for 10 seconds, 
then 

slowly

 exhale directly onto the face of the 

sensor (in the same way you would to fog up a 
piece of glass).  If the descending oxygen alarm is 
set to 19.5%, the instrument should go into alarm 
after a few seconds. 

3.1.2  Verifying accuracy: ToxiPro with 

toxic gas sensor 

Verification of accuracy is a two-step procedure for 
ToxiPro instruments equipped with a toxic gas sensor: 

1. 

Step one is to take the ToxiPro to an area where 
the atmosphere is known to be fresh and check 
the readings.  If the readings differ from those 
expected in fresh air (instruments equipped with a 
toxic sensor should read 0 PPM in fresh air), then 
a fresh air calibration adjustment must be made 
as discussed below in section 3.3. 

2. 

Step two is to test sensor response by exposing 
the sensor to a test gas of known concentration.  
This is known as a functional (bump) test.  
Readings are considered to be accurate when the 
display is between 90% and 120% of the 
expected values as given on the calibration gas 
cylinder.  If readings are accurate, there is no 
need to adjust your gas detector.  See section 3.4 
for further details concerning the functional/bump 
test. 

If the readings are inaccurate, the instrument must be 
span calibrated before further use as discussed in 
section 3.5. 

 

The accuracy of ToxiPro 

instruments equipped with toxic gas sensors should 
be checked periodically with known concentration 
calibration gas.  Failure to check accuracy can lead 
to inaccurate and potentially dangerous readings. 

See Appendix C for a discussion of calibration 
frequency recommendations. 

3.2 

Effect of contaminants on 
ToxiPro sensors 

The atmosphere in which the ToxiPro is used can have 
lasting effects on the sensors.  Sensors may suffer 
losses in sensitivity leading to degraded performance if 
exposed to certain substances. 

The ToxiPro O

2

 uses a galvanic oxygen sensor, while 

toxic sensor versions of the ToxiPro use an 
electrochemical toxic gas sensor.  Different types of 
sensors use different detection principles, so the 
conditions that affect the accuracy of the sensors vary 
from one type of sensor to the next. 

3.2.1  Effects of contaminants on O

2

 

sensors 

Oxygen sensors may be affected by prolonged exposure 
to "acid" gases such as carbon dioxide.  The oxygen 
sensors used in Honeywell Analytics instruments are not 
recommended for continuous use in atmospheres 
containing more than 25% CO

2

3.2.2  Effects of contaminants on toxic 

gas sensors 

Honeywell Analytics’s “substance-specific” 
electrochemical sensors have been carefully designed to 
minimize the effects of common interfering gases.  
“Substance-specific” sensors are designed to respond 
only to the gases that they are supposed to measure.  
The higher the specificity of the sensor, the less likely 
the sensor will react to other gases, which may be 
incidentally present in the environment.  For instance, a 
“substance-specific” carbon monoxide sensor is 
deliberately designed not to respond to other gases that 
may be present at the same time, such as hydrogen 
sulfide (H

2

S) and methane (CH

4

). 

Although great care has been taken to reduce cross-
sensitivity, some interfering gases may still have an 
effect on toxic sensor readings.  In some cases the 
interference may be positive and result in readings that 
are higher than actual.  In other cases the interference 
may be negative and produce readings that are lower 
than actual and may even cause the instrument to 
display negative readings for the target gas. 

See Appendix A for cross-sensitivity data. 

Summary of Contents for ToxiPro CO

Page 1: ...Reference Manual ToxiPro Single Gas Detector Part 13 264 N3 Version 4 ...

Page 2: ...TIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED ToxiPro NGSG Reference Manual Honeywell Analytics Part Number 13 264 N3 Version 4 Copyright 2015 by Honeywell Analytics Inc Lincolnshire Illinois 60069 All rights reserved No page or part of this operation manual may be reproduced in any form without written permission of the copyright owner shown above Honey...

Page 3: ... Missing sensor during startup 9 1 3 6 Corrupt memory 10 1 3 7 Vibrating Alarm Error 10 1 3 8 Heartbeat Feature 10 1 4 Functions 10 1 4 1 Security beep 10 1 4 2 Latching alarms 10 1 4 3 Enhanced Latching Alarm 10 1 4 4 Silence warning alarms 10 1 5 IQ Dock Compatibility 10 1 6 Design components 10 1 7 Standard accessories 11 1 8 Value pack kits 11 2 FIELD OPERATION 11 2 1 Turning the ToxiPro on 11...

Page 4: ...8 3 6 1 2 Forced fresh air zero calibration 18 3 6 2 Span calibration failure 18 3 6 2 1 Sensor out of range no CAL 18 3 6 2 2 Causes for span cal failure 18 3 7 Fresh air zero calibration in a contaminated atmosphere 18 4 MAINTENANCE 19 4 1 Replacing batteries 19 4 2 Replacing sensors 20 4 3 Proper Cleaning 20 4 4 Storage 20 5 PC INSTRUMENT COMMUNICATIONS 20 5 1 Event logging 20 5 2 Instrument fi...

Page 5: ...vision 1 Groups E F G CSA Class I Division 1 Groups A B C D Temp Code T4 CSA Class II Division 1 Groups E F G UL International DEMKO A S 09 ATEX 0823124 II 1 G Ex ia IIC T4 IECEx UL 09 0011 Ex ia IIC T4 Ga KTL 13 KB4BO 0147X http www honeywellanalytics com media honeywell analytics products toxipro documents kcs certificate juarez toxipro 13kb4bo0147x pdf la en The ToxiPro NGSG is in conformity wi...

Page 6: ...otentially dangerous readings The ToxiPro O2 should be periodically calibrated in fresh air A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately An instrument equipped with a toxic gas sensor that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored and the instrument is once again...

Page 7: ...nt à l air libre Tout capteur ne pouvant être étalonné ou hors de sa plage de tolérance doit être remplacé immédiatement Si l étalonnage d un capteur de gaz toxiques n est pas concluant le détecteur concerné ne doit en aucun cas être utilisé Pour que celui ci soit de nouveau utilisable il convient préalablement de le tester en appliquant une concentration connue de gaz de test pour vérifier la pré...

Page 8: ...8 ...

Page 9: ...w In the danger state the LED s will flash red Two oxygen alarm set points have been provided for ToxiPro O2 instruments The danger alarm is sounded for oxygen deficiency and the warning alarm is sounded for oxygen enrichment ToxiPro instruments equipped with a toxic gas sensor have up to four alarm set points Warning Danger Custom TWA Time Weighted Average if enabled and STEL Short Term Exposure ...

Page 10: ...atched When the alarm is latched and if the gas level goes below the alarm level the audible and visual alarms will continued If alarm latching is enabled and the instrument is in alarm for less than the latching duration the alarm will not latch when the gas level drops below the alarm level Latch Acknowledgement A latched alarm latch can be acknowledged by two methods The first is either by mode...

Page 11: ...al the ToxiPro s display will be blank The ToxiPro O2 will show the negated horn icon or To initialize the instrument press the MODE button for 5 seconds ToxiPro O2 models will proceed with a 15 minute countdown while the oxygen sensor stabilizes Note The 15 minute sensor warm up period is only necessary for the initial start up of ToxiPro O2 models When the ToxiPro O2 warm up period concludes the...

Page 12: ...bration may be performed manually or by placing it in an IQ Express Dock Upon successful calibration the ToxiPro will proceed to the current gas readings screen The calibration due lockout function can be enabled or disabled with BioTrak software 2 2 Backlight The ToxiPro includes a backlight that is automatically turned on during an alarm condition The backlight can also be manually activated whi...

Page 13: ...mp test the calibration bottle icon will be displayed at the bottom of the current gas readings screen without the triangular warning symbol Note If an IQ Express Dock is unavailable a span calibration will also reset the bump test due notice 2 5 Low battery alarms When there are less than 7 days of battery life remaining the low battery icon will be lit When there are less than 8 hours of battery...

Page 14: ...aspirated sample draw kit that is available separately During remote sampling the gas sample is drawn into the sensor compartment through the probe assembly and a length of tubing The ToxiPro and manual sample draw kit are delivered with polyester urethane fuel resistant tubing part number 53 001 This material is completely compatible with the toxic gases CO and H2S When using the ToxiPro and samp...

Page 15: ...the calibration gas cylinder If readings are accurate there is no need to adjust your gas detector See section 3 4 for further details concerning the functional bump test If the readings are inaccurate the instrument must be span calibrated before further use as discussed in section 3 5 The accuracy of ToxiPro instruments equipped with toxic gas sensors should be checked periodically with known co...

Page 16: ... lead to inaccurate and potentially dangerous readings 3 4 Functional bump testing toxic sensor versions The accuracy of ToxiPro instruments may be verified at any time by a simple functional bump test To perform a functional bump test do the following 1 Turn the ToxiPro on and wait at least three minutes to allow the readings to fully stabilize If the sensor has just been replaced it must be allo...

Page 17: ...completes the span calibration it will emit three short beeps and display the maximum span calibration adjustment value for two seconds Note The maximum span calibration adjustment value shown is an indication of the relative health of the sensor As a sensor loses sensitivity the maximum adjustment level will approach the calibration gas concentration letting you know when the sensor is losing sen...

Page 18: ...calibration will fail and the no and CAL screens will be alternately displayed After displaying no and CAL times the instrument will return to the current gas readings screen and the warning symbol and the calibration bottle icon will be shown which signifies that the instrument failed the last attempt to span calibrate Note If the Calibration Due Lockout is enabled the ToxiPro will turn itself of...

Page 19: ... and the ToxiPro will show F 0 on the display and will beep 6 times upon startup It will then go into the set time date mode The next five steps describe programming the time and date 7 Press and release the MODE button to advance the setting by one Hold the MODE button down to scroll rapidly Afternoon and evening hours are differentiated by a dot below the colon on the display 10 a m 10 p m Once ...

Page 20: ...he collected data to a PC See section 5 3 for instructions on initiating communications Warning Danger Custom STEL and TWA alarms are logged as events The event logger in ToxiPro version 9 20 or greater logs will nest events The word nested in this case refers to the logging of multiple events simultaneously When the instrument goes from Warning to Danger alarm condition then both Warning and Dang...

Page 21: ...cations To initiate communications hold the MODE button as if you are turning the instrument off Once OFF or on is shown continue to hold MODE until the IrdA screen is shown and the LED IrDA port turns on Once IrdA is shown release the MODE button Then align the infrared port at the top front of the ToxiPro with the PC s infrared port to proceed with the download The LED IrDA port will blink durin...

Page 22: ...lfur Dioxide SO2 1 1 100 1 100 50 150 0 2 n d n d 0 Nitrogen Dioxide NO2 5 8 1 0 100 90 270 0 n d n d 0 n d n d Chlorine Cl2 specific 0 3 0 n d 12 100 20 0 0 0 0 0 0 Chlorine Dioxide ClO2 specific 0 25 0 n d n d 60 100 0 0 0 n d 0 0 Ammonia NH3 54 47 21 0 5 0 n d 0 0 n d 0 0 0 100 0 0 Phosphine PH3 0 5 25 20 n d 0 1 n d n d n d 1 0 5 Hydrogen Cyanide HCN 0 5 350 160 5 100 20 60 0 1 100 65 5 50 n d...

Page 23: ...ytics recommends the following procedure to establish a safe and prudent accuracy check schedule for your Honeywell Analytics instruments 1 During a period of initial use of at least 10 days in the intended atmosphere check the sensor response daily to be sure there is nothing in the atmosphere that is poisoning the sensor s The period of initial use must be of sufficient duration to ensure that t...

Page 24: ...DAMAGE OF ANY KIND CONNECTED TO THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY Instrument Accessory Warranty Periods Product s Warranty Period PHD6 2 years from date of purchase ToxiPro MultiPro 2 years from date of purchase ToxiLtd 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Battery packs and chargers sampling p...

Page 25: ...Honeywell Analytics Inc 405 Barclay Blvd Lincolnshire IL 60069 800 711 6776 Fax 800 995 4992 www honeywellanalytics com ...

Reviews: