6 - 9
Step 6.
Remove the full capacity weight from the scale platter. The green LED blinks 5 times. Wait at least 3 seconds for
scale stability, then press the Sound button once. There is a short delay followed by 5 short beeps that indicates
that the return zero load value has been stored.
Step 7.
The calibration process is complete. The scanner automatically restarts, then beeps once after 12 seconds.
Step 8.
Verify the calibration was successful with the calibrated weights. If the scale display does not show the calibrated
load weight correctly, repeat this procedure.
If you have any issues with the calibration, check that the unit has been properly installed and verify that there is no rocking
motion when pressing on opposite corners of the scale. If you feel any movement, correct by adjusting the leveling screws
that support the scale on the check-stand. When they are correct, lock these down snugly.
Once you have completed the scale calibration, the scale must be verified by the appropriate country or state Weights and
Measures inspector. For details, contact your local government’s Weights and Measures department.
Scale Calibration with Push Buttons
This method of calibration uses the push buttons on the scanner and the LCD display located in the horizontal window of
the scanner scale, visible through the platter.
Step 1.
Press and hold the
Image Capture
and
F1
buttons simultaneously for 5 seconds,
then release them together.
You will see
[MENU]
appear for a second or two on the LCD display, then
SOFT-
WARE UPDATE
. (Do not hold the buttons too long, or you might get a button stuck
error.)
Step 2.
With
SOFTWARE UPDATE
showing on the LCD display, press the
Sound
button
repeatedly until you see
CALIBRATE SCALE
on the LCD display.
Step 3.
Press the
F1
button. The LCD display shows
CALIBRATE SCALE – SINGLE RANGE
KGS
.
Step 4.
Press the
Sound
button repeatedly until the LCD display reads
CALIBRATE SCALE
– SINGLE RANGLE LBS
.
Step 5.
STOP
at this point. Raise the platter and remove the 7mm hex nut and scale service
plate to expose the scale service switch pinhole. Use a thin rod or straightened paper
clip to push down through the exposed pinhole and depress the calibration switch until
a scale beep sounds. The scale is now in service mode. Replace the platter before
continuing.
Scale Zero
F1
Sound
Image Capture
Summary of Contents for Stratos 2700
Page 1: ...Stratos 2700 Bioptic Scanner Scale User s Guide ...
Page 6: ......
Page 20: ...1 6 ...
Page 36: ...2 16 ...
Page 96: ...5 4 ...
Page 110: ...6 14 ...
Page 130: ...8 14 ...
Page 203: ...9 73 Maximum Message Length ...
Page 204: ...9 74 ...
Page 212: ...11 2 Denmark 8 9 6 Norway 8 9 6 Spain 6 Country Codes ...
Page 214: ...11 4 ...
Page 218: ...12 4 ...
Page 244: ...13 26 ...
Page 250: ...14 6 ...
Page 256: ...15 6 ...
Page 270: ...B 2 ...
Page 273: ...Programming Chart 0 1 2 3 4 5 6 7 8 ...
Page 274: ...9 A B C D E F Save Discard ...
Page 276: ......
Page 277: ......