SPM Single Point Monitor
SPM Technical Handbook
4-6
Table 4-1. mA Output for LDL
Gas
mA at LDL
Gas
mA at LDL
DMA Low
4.25
GeH
4
(XPS)
7.76
DMA
5.07
H
2
Se
6.13
MDA
4.82
H
2
Se (XPS)
6.13
TDA
5.07
SbH
3
5.07
TDMAT
5.07
TBA
5.63
HDI
4.25
TBA (XPS)
5.63
N
2
H
4
5.07
TBP
5.63
MMH Low
5.63
TBP (XPS)
5.63
UDMH Low
6.70
H
2
S Low
4.19
AsH
3
5.63
HCI Low
4.38
AsH
3
(XPS)
5.07
HF
5.07
B
2
H
6
5.63
Cl
2
Low (XPS)
5.07
BF
3
Low
5.07
O
3
5.63
Si
2
H
6
5.63
ClO
2
(F
2
/Ox)
5.63
SiH
4
(XPS)
4.34
NH
3
(XPS)
4.31
GeH
4
7.76
SO
2
5.07
4.5 Status Indicators
Condition
Indicator Status
Normal Monitoring
Green system status LED (9) lighted.
Chemcassette
®
Loading
Green system status LED (9) flashing slowly. If
instrument remains in this mode for more than 2
minutes, red system status LED (12) will also begin to
flash, audio alarm pulses and the instrument fault relay
is activated.
Response Verification Green system status LED (9) flashing rapidly. If system
passes test, alarm lamp (16) lights and audio alarm
sounds continuously; if system fails test, red system
status LED (12) lights and audio alarm signals twice.
Instrument
Fault (except
Chemcassette
®
loading fault)
Red and green system status LEDs (9) and (12) are
both flashing. The relay disable LED (13) flashes
(except for the open tape load lever fault), audio alarm
pulses and the instrument fault relay is activated.
Gas Condition
4.6 Display Messages
Status
Display
Normal Monitoring
Actual concentration in ppm (parts-per-million) or ppb
(parts-per-billion). When unit detects an above full-
scale condition, the display will show xxx + ppx; e.g.,
AsH3 above full-scale is 150 + ppb; CI2 reading above
full-scale is 3.0 + ppm.
Verify Mode
(VERIFY)
Instrument Fault
(FAULT) (along with two-digit code; see
Section 4.7
for
codes and corresponding faults or errors)
Alarm Simulation
for displayed information.
4.7 Fault Codes
Instrument Fault or Error:
Message (FAULT) along with two-digit fault code
Fault
Code
Fault or Error
06
RAM Failure, Contact Honeywell Analytics
08
EEPROM Fault, Contact Honeywell Analytics
10
Look-up Table Error, Contact Honeywell Analytics
11
ChemKey Error (See
17
Loss of Flow (See
Numbers 5, 6 & 7)
21
Motor Home Time-out, Contact Honeywell Analytics
25
Gate Fault (Mode Time-out) (See
)
30
High Background Counts (See
No. 3)
32
Reference Voltage #1 Fault, Contact Honeywell Analytics
34
Battery Very Low
36
Over Temp (Heater Option Only) (See
)
Contact Honeywell Analytics if problems persist.
Note:
A fault condition will cancel the alarm relay
disable feature. To resume operations with
alarm relay disabled, you must re-select the
feature. See
Summary of Contents for SPM
Page 8: ...SPM Single Point Monitor Midas Technical Handbook 1 1 1 Operation ...
Page 15: ...SPM Single Point Monitor SPM Technical Handbook 2 1 2 Installation ...
Page 20: ...SPM Single Point Monitor Midas Technical Handbook 3 1 3 Maintenance ...
Page 24: ...SPM Single Point Monitor Midas Technical Handbook 4 1 4 Specifications ...
Page 36: ...SPM Single Point Monitor SPM Technical Handbook 5 1 5 Options ...
Page 53: ...SPM Single Point Monitor SPM Technical Handbook A 1 A Bi Directional Communications Protocol ...
Page 61: ...SPM Single Point Monitor SPM Technical Handbook B 1 B Warranty Statement ...