SPM Single Point Monitor
SPM Technical Handbook
1-2
1.1 Introduction
The SPM combines the well-proven Chemcassette
®
Detection System and microprocessor control to achieve
optimum detection speed, accuracy, and specificity. It
responds quickly to hazardous releases, yet ignores
other commonly used chemicals.
The SPM is very easy to operate and maintain. The
SPM can quickly be set up for monitoring, and routine
maintenance requires less than three hours annually.
These Operating Instructions provide all the information
needed to operate and maintain the SPM Single Point
Monitor. Read them thoroughly before using your new
instrument.
1.2 Master SPM Illustration
The front of this manual contains an illustration of the
SPM together with a numbered legend of all important
parts and controls. Throughout the manual, parts and
controls are identified with the same legend number
used on the illustration.
1.3 Before You Begin
1. Read these Operating Instructions carefully.
2. Attach sample and exhaust tubing. (
3. Connect auxiliary devices such as remote
alarms. (
)
4. Supply power to the instrument. (
)
Warning
When monitoring for certain gases, there may be
limits on the length of sample lines. See
for more information.
Caution
The SPM should not be operated in direct
sunlight or at elevated temperatures unless
equipped with appropriate options. See
for proper operating temperature range.
Caution
Except for instrument maintenance procedures
covered in this manual, all repair and service
work must be referred to qualified personnel.
1.4 Switching Instrument On and Off
To switch the SPM Single Point Monitor on and off,
press the main power switch (8) on the front panel of
the instrument. When the SPM is switched on, the digital
display (19) will scroll the following information:
• Gas Type
• Alarm Levels
• Chemcassette
®
Type
If power is disrupted while the instrument is monitoring,
it will automatically resume monitoring when power is
restored.
Whenever the SPM has been switched off for more than
two hours, advance the Chemcassette
®
tape (7) a few
inches to position fresh tape in the detector head (17).
To manually advance the tape, open the tape load lever
(18) and rotate the take-up reel (27) clockwise.
Note:
For ChemKey operation only—Some newer ChemKeys
will require a Chemcassette
®
that may not be recognized
by the instrument. In these cases, the SPM will display
(Unknown Chemcassette
®
) or (Proper Chemcassette
®
).
However, your ChemKey unit will function correctly
if the correct Chemcassette
®
is installed.
Summary of Contents for SPM
Page 8: ...SPM Single Point Monitor Midas Technical Handbook 1 1 1 Operation ...
Page 15: ...SPM Single Point Monitor SPM Technical Handbook 2 1 2 Installation ...
Page 20: ...SPM Single Point Monitor Midas Technical Handbook 3 1 3 Maintenance ...
Page 24: ...SPM Single Point Monitor Midas Technical Handbook 4 1 4 Specifications ...
Page 36: ...SPM Single Point Monitor SPM Technical Handbook 5 1 5 Options ...
Page 53: ...SPM Single Point Monitor SPM Technical Handbook A 1 A Bi Directional Communications Protocol ...
Page 61: ...SPM Single Point Monitor SPM Technical Handbook B 1 B Warranty Statement ...