GasAlertMicro 5/PID/IR Battery Charger

 [

English] Instruction Sheet

 

D5688/2 Page 

4  

Trimmed and folded dimensions: 4.125 (w) x 5.875 (H) 

 
 

 

5. 

Push the battery pack down, angle it towards the back of the 
charger, and release. Refer to Figure 7. The battery pack sits 
securely if inserted correctly. 

Figure 7. Charging the Battery Pack without the Detector (Side) 

 

The charger emits one beep and the LED lights red. 

6. 

Allow the battery to fully charge (approximately 3.5 hours).

 

The charging LED changes to green when charging is complete. 

When the battery pack is removed from the charger, the LED 
deactivates. 

a

 Warning 

When replacing battery packs, ensure the GasAlertMicro 5 IR battery 
pack is only inserted into the GasAlertMicro 5 IR detector. The IR 
battery packs are not compatible with other Micro 5 models and  
vice versa. Refer to the IR warning label located on the back of the  
IR battery pack lid.  

Troubleshooting 

If a problem occurs, refer to the following solutions. If the problem persists, 
contact 

BW Technologies by Honeywell

Table 3. Troubleshooting Guide

 

Problem 

 

 

Possible 

Cause

Solution

The detector is not 
inserted correctly in the 
cradle. 

Firmly insert the detector 
into the cradle. 

The battery pack is 
above or below the 
operating temperature. 

Allow the battery pack time 
to adjust to the specified 
operating temperature 
(approximately 1 hour). 

The battery pack is 
severely depleted. 

Contact 

BW Technologies 

by Honeywell

The charge 
LED does not 
activate when 
the detector is 
inserted. 

Damaged or defective 
battery pack. 

Contact 

BW Technologies 

by Honeywell

There is no 
audible beep. 

Internal damage to 
battery or charger. 

Contact 

BW Technologies 

by Honeywell

The battery is above or 
below the operating 
temperature. 

Leave the battery in the 
charger. When the battery 
returns to room 
temperature, charging will 
begin. 

The battery pack is 
severely depleted. 

Contact 

BW Technologies 

by Honeywell

There is an 
audible beep, 
but no charge. 

Damaged or defective 
battery pack. 

Contact 

BW Technologies 

by Honeywell

Maintenance 

a

 Warning 

No user-serviceable parts inside. 

To maintain the charger in good operating condition, perform the following 
basic maintenance as required. 

 

Clean the exterior with a soft, damp cloth. Do not use solvents, soaps, 
or polishes. 

 

Do not immerse the charger in liquids. 

Battery Pack Storage 

When stored for extended periods of time, ensure that the battery pack is 
fully charged and then recharged every 30 days. 

Replacement Parts and Accessories 

a

 Warning 

To avoid personal injury or damage to the charger, use only 
the specified replacement parts. 

Table 4. Replacement Parts and Accessories 

Model No. 

Description 

Qty 

M5-BAT01

GasAlertMicro 5 and GasAlertMicro 5 
PID rechargeable battery pack 

M5IR-BAT04

GasAlertMicro 5 IR rechargeable 
battery pack 

M5-C01-BAT01

*

 

GasAlertMicro 5 and GasAlertMicro 5 
& PID cradle charger / battery pack kit 

M5IR-C01-BAT04* 

GasAlertMicro 5 IR cradle charger / 
battery pack kit 

*

Add suffix (-UK) for United Kingdom mains plug, (-EU) for European mains 

plug, (-AU) for Australian mains plug. 

Summary of Contents for M5-C01 Series

Page 1: ...ttery pack with any other unapproved charger Do not use the charger change the batteries or charge the batteries in a hazardous location Failure to observe these precautions will impair the Intrinsic Safety of the detector and can lead to fire and or explosion Do not expose the charger to electrical shock or severe continuous mechanical shock Ensure the battery pack is locked in place prior to ope...

Page 2: ...r emits a low battery alarm charge the battery immediately When charging a new battery for the first time fully charge the battery pack before use approximately 3 5 hours Note If a fully charged battery is inserted into the charger the charger status LED lights red for 6 to 10 minutes and then changes to green does not damage the battery The detector can remain in the cradle after charging is comp...

Page 3: ...the battery pack refer to Figure 5 and complete the following 1 Deactivate the detector 2 Open the latch on the bottom of the detector 3 Remove the battery pack by lifting the bottom of the pack away from the detector 4 Insert the new battery pack Close and press the latch until the release tabs click Figure 5 Removing the Battery Pack Charging the Battery Pack without the Detector To charge the b...

Page 4: ...oneywell The charge LED does not activate when the detector is inserted Damaged or defective battery pack Contact BW Technologies by Honeywell There is no audible beep Internal damage to battery or charger Contact BW Technologies by Honeywell The battery is above or below the operating temperature Leave the battery in the charger When the battery returns to room temperature charging will begin The...

Page 5: ...cts to the original buyer BW s warranty obligation is limited at BW s option to refund of the purchase price repair or replacement of a defective product that is returned to a BW authorized service center within the warranty period In no event shall BW s liability hereunder exceed the purchase price actually paid by the buyer for the Product This warranty does not include a fuses disposable batter...

Page 6: ...ing the details below UK Office Keison Products P O Box 2124 Chelmsford Essex CM1 3UP England Tel 44 0 1245 600560 Fax 44 0 1245 808399 Email sales keison co uk Please note Product designs and specifications are subject to change without notice The user is responsible for determining the suitability of this product ...

Reviews: