50403/SA2/1305
14
HERMetic Sampler A2
6.3 Operating the BOTTOM SAMPLING BOTTLE
(O = OPTION)
ND
TS
DESCRIPTION
O
20246
20124
Bottom bottle 0.40 l FKM assy
1.
Remove the 2” cover from the valve.
2.
Install the sampler on top of the valve.
3. Open the valve.
4.
Lower the bottom bottle until reaching the bottom of the tank.
5.
When the bottom valve of the bottle hits the tank bottom the bottle fills up automatically.
6.
Lift the bottle back into the sampler.
7.
Close the valve.
8.
Remove the cover of the sampler together with the bottom bottle. Mind not to open the
bottom valve inadvertently.
9.
Put the bottom bottle vertically into a laboratory bottle with
a minimum 2” neck diameter.
10. When the bottle bottom valve hits the bottom of the laboratory bottle the liquid is transferred.
11.
Reassemble the bottom bottle and the cover to the sampler.
12. Remove the sampler from the valve.
13.
Reinstall the 2” cover on top of the valve.
14. Clean the equipment after use and check it for proper functioning.