UTImeter Gtex MMC
38
10.3 Battery
replacement
Warning : change the battery only in a non hazardous area.
Unscrew the 2 screws of the battery holder using the 2,5 mm Hex Allen key which is located on the
carrying case. See Figure 10-4.
Pull it gently out.
Change the battery (one-way only device). See Figure 10-4.
Push the battery holder back in its housing (one-way only).
Tighten the 2 screws.
Only one battery is approved:
Duracell / Procell MN1604
Caution:
Do not throw batteries in rubbish; dispose them in a recycling bin.
one-way only device
one-way only device
Figure 10-4