7.3 Tape
The ETFE (TEFZEL) coated tape provides 3 main functions :
• It holds the sensing probe.
• It is graduated and therefore makes it possible to determine
the distance between the reaction point and the reading
index. If the reading index is set up at the zero ullage level,
the reading of the tape is identical to the ullage.
• It contains 2 wires for transmitting the signal and the power
between the display unit and the probe. The steel tape itself
is used as a grounding wire between the sensing probe tube
and the display unit.
The standard graduation is a double side type that shows the metric graduation on one side and the inch one on the other side.
The tape is mounted on the equipment according to the need.
3
4
23
10
11
12
13
14
15
7
7
5
23
9
23
inch side
metric side
Figure 7-7
TEFZEL
STEEL TAPE
WIRE
Figure 7-6