Troubleshooting, Cleaning
27
Error
number
Possible causes
Problem
Display
(flashing)
Err 22
Loop current
- Meas. value above configured current end
- Wrong configuration for current end (see Pg. 20)
Err 23
Loop current
- Configured current span too small
(Difference between current beginning and end)
Err 33
Sensocheck
- Short circuit in primary coil
- Short circuit of cable
Err 34
Sensocheck
- Open circuit in secondary coil
- Cable interrupted
Err 98
System error
- Configuration or calibration data defective; completely
reconfigure and recalibrate the Transmitter
- Measured value transmission defective
- Memory error in Transmitter program (PROM defective)
Err 99
Factory settings
- EEPROM or RAM defective
- Error in factory settings
This error message normally should not occur, as the
data are protected from loss by multiple safety functions.
Should this error message nevertheless occur, there is
no remedy. The Transmitter must be repaired and recali-
brated at the factory.