MAN0604_Issue 09_02-2013
Apex
2110M8030
5. MAINTENANCE
Socket (2 off)
Locating
tab
recess
Sensor body
Plug
(2 off)
Locating
tab
Cartridge
(10) Press
ok
.
The
Change Cartridge
screen now displays the following message.
Change Cartridge
Working
Please Wait
This screen is displayed while the cartridge is checked by the Transmitter Unit.
When the checks are completed successfully the following type of information about the new
cartridge is displayed on the screen.
Change Cartridge
New Cartridge Type
Gas:
HF
FSD:
12.00 ppm
Alarm1:
1.50ppm
Alarm2:
3.00ppm
Check that the information displayed corresponds to the type of replacement cartridge fitted.
(11) Press
ok
.
The
Change Cartridge
screen now displays the following message.
Change Cartridge
Success!
OK: Continue
(12) Press
ok
.
The screen display returns to the
Configuration Menu
.
(13) Press
esc
twice.
The screen returns to the
Gas Reading
display.