background image

INSTALLATION SEQUENCE

PAGE

Step 1. Verify duct air flow direction and velocity ................................................................................................................2
Step 2. Drill mounting holes ................................................................................................................................................2
Step 3. Mount duct housing ................................................................................................................................................3
Step 4. Install inlet tube ......................................................................................................................................................3
Step 5. Complete field wiring ..............................................................................................................................................5
Step 6. Install filters..............................................................................................................................................................7
Step 7. Perform detector check ..........................................................................................................................................7
Step 8. Install cover ............................................................................................................................................................9
Step 9. Perform final system check (Testing) ......................................................................................................................9

STEP 1. VERIFY DUCT AIR FLOW DIRECTION AND VELOCITY

The 14506873-001 is designed to be used in air handling systems having air velocities of 500 to 4000 ft  (152.5 to 1219
m) per minute. Be sure to check engineering specifications to ensure that the air velocity in the duct falls within these
parameters. If necessary, use a velocity meter to check the air velocity in the duct.

STEP 2. DRILL MOUNTING HOLES

Remove the paper backing from the drilling template. Affix the template to the air duct at the desired mounting location.
Make sure the template lies flat and smooth on the air duct. Center punch hole targets and remove the template. Drill the
holes as indicated on the template. Slide the two speed nuts over the two small holes (Hole A) next to the sampling tube
bushing holes (Hole B) previously drilled in the duct. (See Figure 2.)

Location of detectors mounted in or on air ducts should be at least six duct widths downstream from any duct openings,
deflection plates, sharp bends, or branch connections.

Exception: Where it is physically impossible to locate the detector accordingly, the detector can be positioned closer than
six duct widths, but as far as possible from the opening, bend, or deflection plates.

STEP 3. MOUNT DUCT HOUSING

Remove the duct housing cover. Slide the foam gaskets over the tube bushings as shown in Figure 2A. Make sure the
two small holes in the gaskets line up with the two duct housing mounting holes. Put one 5/16-inch O-ring over each of
the two #10 sheet metal screws. Use the two sheet metal screws to secure the duct housing to the duct.

CAUTION: Do not overtighten the screws.

A78-2046-00

A78-2449-00

Figure 2. Speed Nut Mounting Location.

Figure 2A. Installation of Foam Gaskets Over

Sampling Tube Bushings.

TABLE 1. SAMPLING (INLET) TUBES

TUBE

OUTSIDE DUCT WIDTH

SSDST-1.5

1 to 2 ft.

(30 to 61 cm)

SSDST-3

2 to 4 ft.

(61 to 122 cm)

SSDST-5

4 to 8 ft.

(1.2 to 2.4 m)

SSDST-10

8 to 12 ft.

(2.4 to 3.7 m)

SCREW HOLES FOR ATTACHING
DETECTOR HOUSING TO DUCT

FOAM GASKETS

EXHAUST TUBE

(EXTENSION BUSHING)

DUCT DETECTOR

HOUSING

O-RINGS

MOUNTING

SCREWS

INLET SAMPLING

TUBE BUSHING

HOLE B

HOLE A

DUCT
WALL

SPEED

NUT

A78-2454-00

Figure 10. Photo Head Exploded View

REMOVABLE
COVER

REPLACEABLE
SCREEN

COVER
REMOVAL
SLOT

VANED
SENSING
CHAMBER

PHOTO UNITS
a. Remove detector cover by inserting a small-bladed screwdriv-

er into the cover removal slot located approximately 90
degrees from the test module socket, twisting cover counter-
clockwise to remove. See Figure 10.

b. Lift the screen from the photo chamber. Vacuum the screen

and cover before using clean, compressed air to loosen and
blow out any remaining debris. (Replacement screens, part no.
SSDRS24, are available.)

c. Vacuum photo chamber, then use clean compressed air to

blow area clean.

d. Replace screen by aligning arrow on top with the field test slot

on the base of the detector. Push screen into place. Screen
should fit tightly to chamber.

ION UNITS
a. Remove the detector cover and screen assembly by depress-

ing the three lock prongs on the top of the cover and rotating
the cover counterclockwise. The SSDCRT400 Cover Removal
Tool makes cover removal easier. See Figure 11.

b. Carefully pull the screen out of the cover.
c. Clean the screen thoroughly with a soft brush or vacuum.

(Replacement screens, part no. SSDRS14.)

d. Brush or vacuum the inside of the cover. Cover may then be

blown out using clean, compressed air.

e. Vacuum sensing chamber before using clean, compressed air

to loosen and blow out any remaining debris.

f.

Press the screen back into the cover.

g. Replace the detector cover on the sensing chamber. Rotate it

clockwise to lock it into place.

BOTH TYPES OF UNITS
5. Reinstall the detector in its housing.
6. Restore system power.
7. Put detector into alarm using appropriate method as described

in STEP 7. PERFORM DETECTOR CHECK.

8. Notify the proper authorities that testing has been completed

and the smoke detector system is back in service.

9. Other checks that should be made during maintenance proce-

dures:

— Holes or cracks in duct work near vicinity of detector.
— Air leaks where detector housing or sampling tubes are

attached to duct.

— Dust accumulations in or on sampling tubes.
— Screw terminal tightness.

SPECIFICATIONS

Length:

14.5 inches (36.7 cm.)

Width:

5 inches (12.7 cm.)

Depth:

4 inches (10.2 cm.)

Weight:

4 lbs. (1.8 kg.)

Operating Temperature Range:

32° to 120°F (0° to 49°C)

Operating Humidity Range:

10% to 93% Relative Humidity

Duct Air Velocity:

500 – 4000 ft./min. (157.5 – 1219.2 m/min.)

A78-2455-00

Figure 11. Ion Head Exploded View

LOCK PRONG

REMOVABLE SCREEN

REMOVABLE 

COVER

FOR

CLEANING 

SENSING CHAMBER

H500-03-00

10

I56-475-09R

H500-03-00

3

I56-475-09R

Summary of Contents for 14506873-001

Page 1: ...72 requirements The detector should be cleaned at least once a year TABLE OF CONTENTS GENERAL DESCRIPTION 1 Contents of the Duct Detector Housing Kit 2 INSTALLATION SEQUENCE 3 Verify Duct Air Velocity 3 Drill the Mounting Holes 3 Mount Duct Housing 3 Install the Inlet Tube 4 Ducts Less Than 8 ft 244 cm Wide 4 Ducts More Than 8 ft 244 cm Wide 5 Interim Modification of Inlet Tube 5 Field Wiring 5 Wi...

Page 2: ...respective code commands from the panel See panel instructions for details CONTENTS OF THE DUCT DETECTOR HOUSING KIT The 14506873 001 Air Duct Detector Housings consist of the following items See Figure 1 See Note 1 for Inlet Sampling Tube and Detector Supply information 1 Complete housing base and cover assembly 2 Two 10 sheet metal mounting screws 3 Two sampling tube filters 4 One test magnet 5 ...

Page 3: ...he duct See Figure 2 Location of detectors mounted in or on air ducts should be at least six duct widths downstream from any duct openings deflection plates sharp bends or branch connections Exception Where it is physically impossible to locate the detector accordingly the detector can be positioned closer than six duct widths but as far as possible from the opening bend or deflection plates STEP ...

Page 4: ...t tube If the inlet tube is shorter than the width of the air duct install the end cap into the inlet tube as shown in Figure 3 2 Slide the inlet tube into the bushing located in the center of the duct housing Position the tube so that the arrows point into the air flow 3 Secure the tube flange to the housing bushing with the two 6 self tapping screws 4 For tubes longer than the width of the air d...

Page 5: ...ibration This vibration can slowly open the seal around the tube and permit air to escape To prevent this from occurring a 3 inch floor flange available at most plumbing supply houses may be used This flange connector mounting technique makes the initial installation easier because a 1 in to 1 1 4 in 25 to 32 mm hole may be drilled where the flange connector will be used It is easier to push the i...

Page 6: ...CCESSORIES MAY BE USED SUCH THAT THE GIVEN CURRENT LOADS TOTAL AUX POWER TEST STANDBY STATE ALARM STATE 26mA MAX AT 24 VDC 87mA MAX 24 VDC TERMINALS 3 6 20mA MAX 12 VDC ALARM CONTACTS TERMINALS 2 6 NOTE THE SSDRTS451 SSDRTS451KEY RESET CANNOT BE USED ON THE 14506873 001 TEST RESET NOT USED 4 5 6 7 8 9 10 11 12 13 14 15 16 17 COMMUNICATION LINE FROM PANEL COMMUNICATION LINE TO REST OF SYSTEM NOTE F...

Page 7: ...ust accumulation System filters remove a large percentage of this contami nation but cannot remove all of it Dust inside the duct detector causes problems First very fine particles of dust can enter the detector sensing chamber and cause the unit to go into alarm Second the accumulation of dust and dirt necessitates a more frequent periodic cleaning schedule which can result in substantial cost an...

Page 8: ...utside of the housing see Figure 7 3 The LEDs on the detector should latch on as should any accessories SSDPA400 SSDRA400Z SSDRTS451 and the alarm condition should be verified at the control panel The control panel must be active for an alarm to be ini tiated A78 2452 00 A78 2453 00 Figure 7 Testing Detector Alarm Figure 8 Detector Head Removal B SSDRTS451 SSDRTS451KEY REMOTE TEST STATION The SSDR...

Page 9: ...nel should indicate an alarm condition PERIODIC MAINTENANCE REQUIREMENTS Air duct smoke detectors should be maintained at least once a year more often if the detector heads become obviously dirty The detectors must also be cleaned immediately after a fire Failure to properly maintain air duct smoke detectors may cause unnecessary false alarms Honeywell recommends that a permanent Detector Test Log...

Page 10: ... screens part no SSDRS14 d Brush or vacuum the inside of the cover Cover may then be blown out using clean compressed air e Vacuum sensing chamber before using clean compressed air to loosen and blow out any remaining debris f Press the screen back into the cover g Replace the detector cover on the sensing chamber Rotate it clockwise to lock it into place BOTH TYPES OF UNITS 5 Reinstall the detect...

Page 11: ...806A1037 or TC806A1052 are used the constraints above do not apply 2 The duct control circuitry auxiliary relay contacts and auxiliary alarm output may change states if the sensor is test ed and the LED is turned on for more than 4 seconds If the LED is latched on for less than 4 seconds the control circuitry of the 14506873 001 will NOT operate 3 The panel must not continuously poll the same unit...

Page 12: ...__________ ________ __________ ________ ______________ ____________________________ ________ __________ ________ ______________ ____________________________ ________ __________ ________ ______________ ____________________________ ________ __________ ________ ______________ ____________________________ ________ __________ ________ ______________ ____________________________ ________ __________ ____...

Reviews: