Guidance and manufacture’s declaration – electromagnetic immunity
The CIRCULATION PRO is intended for use in the electromagnetic environment specified below. The
customer or the user should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 V/ms
150 kHz to 80 MHz
3 V/m
26 MHz to 2.5 GHz
10 V/m
26 MHz to 2.5 GHz
3 V/ms
3 V/m
10 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the ELECTRO FLEX, including cables,
than the recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
d = 1,2
√—
P
d = 1,2
√—
P
80 MHz to 800 MHz
d = 2,3
√—
P
800 MHz to 2.5 GHz
where
P
is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter
manufacturer and
d
is the recommended
separation distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site
survey, a should be less than the compliance
level in each frequency range b.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy.
a
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the CIRCULATION PRO is
used exceeds the applicable RF compliance level above, the CIRCULATION PRO should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as reorienting or relocating the CIRCULATION PRO.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
IB-CB200EU-0613-01_Layout 1 26/06/2013 16:06 Page 22
Summary of Contents for Circulation Pro CB-200-EU
Page 11: ...11 GB IB CB200EU 0613 01_Layout 1 26 06 2013 16 06 Page 11 ...
Page 12: ...12 GB IB CB200EU 0613 01_Layout 1 26 06 2013 16 06 Page 12 ...
Page 13: ...13 GB IB CB200EU 0613 01_Layout 1 26 06 2013 16 06 Page 13 ...
Page 33: ...33 F IB CB200EU 0613 01_Layout 1 26 06 2013 16 06 Page 33 ...
Page 34: ...34 F IB CB200EU 0613 01_Layout 1 26 06 2013 16 06 Page 34 ...
Page 35: ...35 F IB CB200EU 0613 01_Layout 1 26 06 2013 16 06 Page 35 ...
Page 55: ...55 D IB CB200EU 0613 01_Layout 1 26 06 2013 16 06 Page 55 ...
Page 56: ...56 D IB CB200EU 0613 01_Layout 1 26 06 2013 16 06 Page 56 ...
Page 57: ...57 D IB CB200EU 0613 01_Layout 1 26 06 2013 16 06 Page 57 ...
Page 77: ...77 ES IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 77 ...
Page 78: ...78 ES IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 78 ...
Page 79: ...79 ES IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 79 ...
Page 99: ...99 I IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 99 ...
Page 100: ...100 I IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 100 ...
Page 101: ...101 I IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 101 ...
Page 121: ...121 P IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 121 ...
Page 122: ...122 P IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 122 ...
Page 123: ...123 P IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 123 ...
Page 143: ...143 NL IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 143 ...
Page 144: ...144 NL IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 144 ...
Page 145: ...145 NL IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 145 ...
Page 165: ...165 TR IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 165 ...
Page 166: ...166 TR IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 166 ...
Page 167: ...167 TR IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 167 ...
Page 187: ...187 EL IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 187 ...
Page 188: ...188 EL IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 188 ...
Page 189: ...189 EL IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 189 ...
Page 209: ...209 RU IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 209 ...
Page 210: ...210 RU IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 210 ...
Page 211: ...211 RU IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 211 ...
Page 231: ...231 PL IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 231 ...
Page 232: ...232 PL IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 232 ...
Page 233: ...233 PL IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 233 ...
Page 253: ...253 S IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 253 ...
Page 254: ...254 S IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 254 ...
Page 255: ...255 S IB CB200EU 0613 01_Layout 1 26 06 2013 16 07 Page 255 ...
Page 275: ...275 DK IB CB200EU 0613 01_Layout 1 26 06 2013 16 08 Page 275 ...
Page 276: ...276 DK IB CB200EU 0613 01_Layout 1 26 06 2013 16 08 Page 276 ...
Page 277: ...277 DK IB CB200EU 0613 01_Layout 1 26 06 2013 16 08 Page 277 ...
Page 297: ...297 N IB CB200EU 0613 01_Layout 1 26 06 2013 16 08 Page 297 ...
Page 298: ...298 N IB CB200EU 0613 01_Layout 1 26 06 2013 16 08 Page 298 ...
Page 299: ...299 N IB CB200EU 0613 01_Layout 1 26 06 2013 16 08 Page 299 ...
Page 319: ...319 FIN IB CB200EU 0613 01_Layout 1 26 06 2013 16 08 Page 319 ...
Page 320: ...320 FIN IB CB200EU 0613 01_Layout 1 26 06 2013 16 08 Page 320 ...
Page 321: ...321 FIN IB CB200EU 0613 01_Layout 1 26 06 2013 16 08 Page 321 ...
Page 341: ...341 CZ IB CB200EU 0613 01_Layout 1 26 06 2013 16 08 Page 341 ...
Page 342: ...342 CZ IB CB200EU 0613 01_Layout 1 26 06 2013 16 08 Page 342 ...
Page 343: ...343 CZ IB CB200EU 0613 01_Layout 1 26 06 2013 16 08 Page 343 ...
Page 363: ...363 HU IB CB200EU 0613 01_Layout 1 26 06 2013 16 09 Page 363 ...
Page 364: ...364 HU IB CB200EU 0613 01_Layout 1 26 06 2013 16 09 Page 364 ...
Page 365: ...365 HU IB CB200EU 0613 01_Layout 1 26 06 2013 16 09 Page 365 ...
Page 385: ...385 SK IB CB200EU 0613 01_Layout 1 26 06 2013 16 09 Page 385 ...
Page 386: ...386 SK IB CB200EU 0613 01_Layout 1 26 06 2013 16 09 Page 386 ...
Page 387: ...387 SK IB CB200EU 0613 01_Layout 1 26 06 2013 16 09 Page 387 ...
Page 398: ...IB CB200EU 0613 01_Layout 1 26 06 2013 16 09 Page 398 ...
Page 399: ...IB CB200EU 0613 01_Layout 1 26 06 2013 16 09 Page 399 ...