Guidance and manufacture’s declaration – electromagnetic immunity
the BPA-3000-EU is intended for use in the electromagnetic environment specified below. the customer of
the user of the BPA-3000-EU should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment - guidance
Electrostatic
discharge (Esd)
iEc 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or ceramic
tile. if floor are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
iEc 61000-4-4
±2 kV for power
supply lines
±2kV for power
supply lines
Mains power quality should be that of a
typical commercial or hospital environment
surge iEc
61000-4-5
±1 kV line(s) to
line(s)
±1 kV differential
mode
Mains power quality should be that of a
typical commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines iEc
61000-4-11
<5% Ut
(>95% dip in Ut)
for 0.5 cycle
<5% Ut
(>95% dip in Ut)
for 0.5 cycle
Mains power quality should be that of a
typical commercial or hospital environment.
if the user of the BPA-3000-EU requires
continued operation during power mains
interruptions, it is recommended that the
BPA-3000-EU be powered from an
uninterruptible power supply or a battery.
40% Ut
(60% dip in Ut)
for 5 cycles
40% Ut)
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
<5% Ut
(>95% dip in Ut)
for 5 sec
<5% Ut
(>95% dip in Ut)
for 5 sec
Power frequency
(50Hz) magnetic
field iEc
61000-4-8
3A/m
3A/m
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
notE: Ut is the a.c. mains voltage prior to application of the test level.
table 2 Guidance and manufacture’s declaration – electromagnetic immunity – for all