6.
Wait until Drop Symbol appears in
Display. Keep Strip in Meter until
testing is finished.
Note:
If Strip is removed before testing
is finished, an error message
appears. Release and discard
old Strip. Use new Strip for
testing.
7.
Turn Control bottle upside down.
Squeeze one drop of Control onto
a clean tissue. Wipe off bottle tip.
8.
Gently squeeze a Control drop onto
a small piece of unused aluminum
foil or clear plastic wrap.
9.
With Strip still in Meter, touch edge
of Sample Tip to drop of Control.
Allow drop to be drawn into Strip.
10.
Dashes appear across the Display to
show Meter is testing. Remove Strip
from drop.
Note:
If Meter does not begin testing soon after
drawing up sample, release and discard
Strip. Repeat test with new Strip. If problem
persists, see
Troubleshooting
.
17
Drop Symbol
Meter Testing
T2go_OB_GutsENG_F4HDI04.qxd:TT_OB_GutsENG.qxd 1/29/10 10:39 AM Page 17