12
HomeBrace Germany UG
(haftungsbeschränkt)
Konrad-Hornschuch-Str. 67 | Halle 24, C1 | 73660 Urbach
+49 7181 255700 | [email protected]
www.homebrace.com
Time
The Time button indicates the time in seconds left or available for the
selected switches�
Example:
Sw1 and T1 selected: keep eyes closed for one second
Example:
Sw2 and T3 selected: blink once (close and open your eyes) within 3
seconds while focusing on the selected button�
Recommended procedure: The sw2 and T3 setting is recommended
for switching modes.
15-02 Calibration
Calibration is required before using MyEcc Pupil for control� In the menu, activate the Calibrate button�
Calibrate Ready
appears�
Start looking around and focusing in all directions� A sound and a blue dot indicate the first calibration run� A green dot
appears, the sound changes and the second stage of calibration starts�
Continue looking around� The sound of a beep indicates that calibration is complete� The system is now ready for activati-
on� The calibration can be saved with the button “Save calibration” and after a restart be called again�
15-03 Activation
Eye focus must alternate between left – right – left� The feedback is a beep and all buttons green (=activated)�
15-04 Connecting to the wheelchair
Functions 1 – 5 (drive forward / right / back / left + function 5) are on a 9-pin D-sub connector as indicated in R-Net
I/O module SK78814/1 wiring diagram� (D-sub pins 1, 2, 3, 4, 6, common: pin 8)� Functions 6 / 7 / 8 are available on
separate jacks�
1 = forward
2 = right
3 = back
4 = left
5 = button center top + switch command
6 = button center right + switch command
7 = button center switch command
8 = button center left + switch command