Electronic Digital Thermometer Operation Instructions MT-B231
FEATURES
1. Internal function test will be carried out every time when the unit is turned on. It is so-called automatic self-test function. If it displays
, it means the thermometer is not
working properly. Please contact the supplier.
2. Temperatures can be taken from under an arm (external use) or under the tongue (oral use). It is suggested that rectal temperatures are taken on infants only when
external use is inconvenient.
3. This is a very sensitive device and results can be obtained within 60 ~ 90 seconds approximately.
4. If the thermometer is inadvertently left on after reading your temperature, it will automatically shut off after 10 minutes.
5. This is a small and light-weight unit (10 grams) and can be used by the whole family.
6. The ABS resin body presents no hazards compared to a broken mercury glass thermometer.
PARTS DESCRIPTION
TECHNICAL SPECIFICATIONS
Measurement range:
89.60 °F ~ 107.60 °F
Accuracy:
Fahrenheit:
±0.30 °F Less than 95.90 °F
±0.20 °F 95.90 °F ~ 107.60 °F
±0.30 °F 107.60 °F ~ 109.40 °F
Display unit:
0.10 °F
Battery:
One 1.55V button size battery (LR 41)
Power consumption:
0.15 MW in measurement mode
Battery life:
More than 200 hours of continuous operation
Weight:
Approx. 10 grams including battery
Dimensions:
132.3mm x 20mm x 14mm
Storage temperature:
-13 °F ~ 131 °F
Operation temperature:
50 °F ~ 95 °F
Max. Storage Humidity:
80% ~ 90%
Type:
Non-predictive
The digital thermometer is waterproof.
HOW TO REPLACE THE BATTERY
When the battery is almost exhausted, the battery symbol “
" will appear on the LCD.
● Slide back the battery cover.
● Use a pointed object to remove the battery from its holder carefully.
● Insert a new battery (LR41) with the “+” symbol at the top.
● when replacing the battery cover, take care that the seal is not damaged.
NOTE:
Batteries should not be disposed of with household waste. Please follow local battery disposal regulations.
WARNING
1. Do not swallow the battery which may be fatal.
2. Do not recharge the battery.
3. Do not dispose of the battery in fire as it may explode.
4. Do not drop the unit.
5. Do not boil the sensor, instead clean the unit by wiping it with dry cloth and clean the sensor with medical alcohol (95%).
6. Do not expose the unit to direct sunlight or high temperatures.
7. Do not bend the unit out of shape.
8. Do not attempt to disassemble the unit except battery replacement.
9. Do not bite or pull the unit.
10. When inserting the unit (for oral and rectal use), insert it gently. Do not insert it strongly or bend it.
HOW TO OPERATE
1.
Press button
, and the LCD screen ○ will display 188.88 for approximately 2 seconds. At the same time, an “LCD test” (to check whether all segments of the display
appear) and an internal function test are carried out.
1.
After switching on the thermometer the display will show the last memorized data for approx. 2 seconds.
2.
The indicator will then display “Lo” for about 0.6 seconds.
3.
While the temperature is being measured, the current temperature will be permanently indicated and the “°F” mark will flicker during the measuring time.
4.
When the fluctuation of the temperature is below 0.04°F for 8 seconds, the “°F” mark will stop flickering and an electronic buzzer will sound. At the same time the final
temperature will be shown on the LCD. From now on the indicated temperature will fluctuate minimally.
5.
After measuring your temperature the temperature will be displayed on the LCD screen until you switch off the thermometer.
6.
To prolong the life of the battery, please switch off the thermometer after use by pressing button
. If the power is left on, the thermometer will automatically shut down
after 10 minutes to save power.
NOTE1:
When measuring the temperature under your arm, it is recommended that keeping the thermometer in position for 1 to 2 minutes. The buzzer will sound several
times during measuring time.
NOTE2:
When the body temperature reaches 100.04°F, the fever alarm will sound.