SMI-1414-EN Rev. 005 - Installation and Service Manual
Page 39 of 51
MCQA Test Image
Associated Coil
GEHC P/N
SMI P/N
Test#=1 SIG_IMG=1 Left Single Loop
5407765-2
4000452-11
Test#=1 SIG_IMG=2 Medial Array
5407765-6
4000445-11
Test#=1 SIG_IMG=3 Medial Array
5407765-6
4000445-11
Test#=1 SIG_IMG=4 Medial Array
5407765-6
4000445-11
Test#=1 SIG_IMG=5 Medial Array
5407765-6
4000445-11
Test#=2 SIG_IMG=1 Medial Array
5407765-6
4000445-11
Test#=2 SIG_IMG=2 Medial Array
5407765-6
4000445-11
Test#=2 SIG_IMG=3 Medial Array
5407765-6
4000445-11
Test#=2 SIG_IMG=4 Medial Array
5407765-6
4000445-11
Test#=2 SIG_IMG=5 Right Single Loop
5407765-3
4000448-11
Additional information related to the specific channels used in each imaging configuration is available in
section 8.1 Imaging Troubleshooting.
7.5.3 Troubleshooting tips
1. Ensure that no patient has been started before beginning the MCQA procedure.
2. Landmark the coils before starting the MCQA tool.
3. Only use the spherical GE phantoms (GE Model 2360034), and the Sentinelle phantom
positioners.
4. Ensure that all coils are touching the phantoms.
5. When the MCQA tools starts, you may see the message shown in Figure 35. If so, click ‘Yes’.
Figure 35: MCQA Warning Message
7.6 Final Verification Checklist
Final verification of install is an important series of functional and visual checks which are undertaken to
ensure that the installation procedure was executed correctly.
SMI
ONLY
Complete the Installation Verification Checklist (section 14 Appendix A) and submit to SMI
Service department.
GEHC
ONLY
Complete the Installation Verification Checklist (section 14 Appendix A) and store on site.