4.2. Common Questions
), scan and process the requested views, and send the
results to the CAD display unit. Then, turn Archiving back on, scan and process all
films for the case, and send the results to the PACS.
Scanned case before entering patient information
You can enter the patient information after scanning and reprocess the study to link
the images to the proper patient. You must include the barcode from the separator
sheet.
•
See ‘
Part 7: Using DigitalNow and the R2 Patient ID System
’ for more
information on entering patient information.
•
See ‘
3.10. Correcting Scanned Images
’ for information on reprocessing a study.
Yellow ‘X’ appears on my image thumbnails
The yellow ‘X’ icon indicates that CAD processing could not be performed on an
image.
•
Verify that the images have the correct view labels. If they do not, correct the view
labels on the Image Correction screen. See ‘
3.10. Correcting Scanned Images
more information.
•
Verify that the image views and view modifiers are supported for CAD processing.
See ‘
2.6. System Inputs and Outputs
’ for more information.
Even though no processing was possible, the scanned images are archived as usual.
R2 DM User Manual – PN MAN-01044 Rev 001
59
Summary of Contents for R2 DM
Page 1: ...User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001 R2 DM...
Page 2: ......
Page 3: ...Screen Film Mammography R2 DM User Manual Software Versions 8 6 8 7 PN MAN 01044 Rev 001...
Page 56: ...Part 3 Studies 48 R2 DM User Manual PN MAN 01044 Rev 001...
Page 68: ...Part 4 Alerts 60 R2 DM User Manual PN MAN 01044 Rev 001...
Page 92: ...Part 6 Maintenance 84 R2 DM User Manual PN MAN 01044 Rev 001...
Page 100: ...92 R2 DM User Manual PN MAN 01044 Rev 001...
Page 105: ......
Page 106: ......