NovaSure Disposable Device Description 

The NovaSure disposable device consists of a single-patient use, 
conformable bipolar electrode array mounted on an expandable 
frame that can create a confluent lesion on the entire interior 
surface area of the uterine cavity . The disposable device is inserted 
transcervically into the uterine cavity, and the sheath is retracted to 
allow the bipolar electrode array to be deployed and conform to the 
uterine cavity . 
The bipolar electrode array is formed from a metalized, porous 
fabric through which steam and moisture are continuously 
suctioned from the desiccated tissue . The disposable device 
works in conjunction with a dedicated NovaSure RF controller to 
perform customized, global endometrial ablation in an average 
of approximately 90 seconds without the need for concomitant 
hysteroscopic visualization or endometrial pretreatment . The 
specific configuration of the bipolar electrode array and the 
predetermined power of the controller create a controlled depth of 
ablation in uteri sounding less than or equal to 10 cm and having a 
minimum cornu-to-cornu distance of 2 .5 cm .
During the ablation process, the flow of radio frequency (RF) energy 
vaporizes and/or coagulates the endometrium regardless of its 
thickness and desiccates and coagulates the underlying, superficial 
myometrium .
The controller automatically calculates the optimal power level (W) 
required for the treatment of the uterine cavity, based on uterine 
size . As tissue destruction reaches an optimal depth, increasing 
tissue impedance causes the controller to automatically terminate 
power delivery, thereby providing a self-regulating process . Blood, 
saline and other liquid present in the uterine cavity at the time 
of the procedure, as well as vapor liberated from the desiccated 
tissue, are evacuated by continuous, automatic suctioning .
The disposable device is connected to the controller via a 
cord containing the RF cable, suction tubing used for pressure 
monitoring during the cavity integrity assessment cycle and for 
suction during the ablation cycle, and vacuum feedback tubing used 
for carbon dioxide delivery during the cavity integrity assessment 
cycle and vacuum monitoring during the ablation cycle . 

NovaSure RF Controller Description 

The NovaSure RF controller is a constant power output generator with a 
nominal maximum power delivery capability of 180 watts . The controller 
automatically calculates the power output based on the uterine cavity 
length (sound measurement minus the length of the endocervical canal) 
and width measurements that the user key-enters into the controller . 
Monitoring tissue impedance during the ablation process automatically 
controls the depth of endo-myometrial ablation . The NovaSure procedure 
self-terminates once endometrial vaporization and superficial myometrial 
desiccation have reached 50 ohms of impedance at the tissue-electrode 
interface, or when the treatment timer reaches two minutes . Integral 
to the controller is the cavity integrity assessment system (CIA) which 
is designed to determine whether there is a defect or perforation in the 
wall of the uterus . After the disposable device is placed into the uterine 
cavity, CO

2

 is delivered through the central lumen of the disposable 

device into the cavity, via the vacuum feedback tubing, at a safe flow 
rate and pressure . If the CO

2

 pressure in the cavity is maintained for a 

short period of time, indicating that the uterine cavity is intact, then the 
CIA will allow the NovaSure RF controller to be enabled and proceed with 
the treatment phase . A vacuum pump contained within the NovaSure 
RF controller creates and maintains a vacuum in the uterine cavity 
throughout the endometrial ablation procedure . Once the vacuum is 
stabilized, the vacuum level is monitored throughout the remainder of 
the ablation process .

NovaSure Model 08-09  

RF Controllers

NovaSure Model 10  

RF Controller

Summary of Contents for NovaSure Advanced

Page 1: ......

Page 2: ...d the disposable device may contain a more recent revision of the NovaSure system instructions than the manual provided with the controller The NovaSure disposable device is not to be used with other...

Page 3: ...suctioning The disposable device is connected to the controller via a cord containing the RF cable suction tubing used for pressure monitoring during the cavity integrity assessment cycle and for suct...

Page 4: ...hyperplasia a patient with any anatomic condition e g history of previous classical cesarean section or transmural myomectomy or pathologic condition e g long term medical therapy that could lead to...

Page 5: ...the disposable device external to the patient to eliminate the risk of air or gas embolism The NovaSure RF controller CAVITY ASSESSMENT LED flashes red Model 08 09 RFCs or a purging device screen app...

Page 6: ...after inserting the disposable device and proceed directly from the seating procedure to the CIA Electrically conductive objects e g monitoring electrodes from other devices that are in direct contact...

Page 7: ...e g hormone D C or patient timing Patients randomized into the control arm received wire loop resection as an endometrial pretreatment Study endpoints The primary effectiveness measure was a validated...

Page 8: ...rameters between the treatment groups between the age groupings or among the nine investigational sites Table 2 Patient Accountability Number of Patients NovaSure Wire Loop Resection Plus Rollerball E...

Page 9: ...culty Performing Work Or Other Activities Due to Menses Pre operatively 66 3 65 5 Table 4 Effectiveness Quality of Life QOL NovaSure Wire Loop Resection Plus Rollerball 12 Months 9 9 8 6 24 Months 14...

Page 10: ...use only in women who do not desire to bear children because the likelihood of pregnancy is significantly decreased following the procedure Patients of childbearing capacity should be cautioned of pot...

Page 11: ...5 Press the toggle switch on the back panel of the controller into the on position 1 6 Connect the foot switch to the appropriate port on the front panel of the controller NOTE The first time the Mode...

Page 12: ...ay can be extended CONTRAINDICATION Do not treat a patient with a uterine cavity length that is less than 4 cm as cervical canal damage may occur NOTE Patients with a uterine cavity length greater tha...

Page 13: ...2 18 Unlock the disposable device by pressing the lock release button Close the disposable device by holding the front handle stationary and gently pulling the rear handle backwards until the closed...

Page 14: ...eeze the disposable device handles together while gently moving the disposable device 0 5 cm to and from the fundus and rotating the handle of the disposable device 45 counterclockwise from the vertic...

Page 15: ...m 2 34 Select the value indicated on the WIDTH dial into the NovaSure RF controller input screen by depressing the UP DOWN arrows 2 35 The system can be operated in either automatic mode or manual mod...

Page 16: ...ting 24 Replacement Instructions 24 Specifications 24 Cleaning and Sanitizing 28 Parts List 28 Warranty 28 Technical Support and Product Return Information 28 Symbol Definitions 29 MODEL 08 09 RF CONT...

Page 17: ...o be at the cervix and cannot be resolved by using the cervical collar use another tenaculum to grasp the cervix around the sheath Repeat the CIA by pressing the foot switch NOTE CO2 leakage may occur...

Page 18: ...modification of this equipment is allowed Periodic maintenance Recommended periodic maintenance includes RF power output testing and standard safety tests such as leakage current in accordance with UL...

Page 19: ...ated when the array is not fully deployed The ENABLE LED cannot be toggled on nor can power be delivered to the array when the ARRAY POSITION LED is illuminated VACUUM LED illuminates in two condition...

Page 20: ...e new device abort the procedure NOTE Removing the disposable device from the uterine cavity after completing a cavity integrity assessment will require an additional CIA test to be performed upon dis...

Page 21: ...2 leakage may occur at the external cervical os due to the presence of an over dilated cervix Visible bubbles or the hissing sound of escaping gas may accompany CO2 leakage under either of these condi...

Page 22: ...ng and standard safety tests such as leakage current in accordance with UL 60601 1 along with regular cleanings described in the Cleaning and Sanitizing section RF power output test The NovaSure RF co...

Page 23: ...r dilated cervix Visible bubbles or the hissing sound of escaping gas may accompany CO2 leakage under either of these conditions NOTE When following the troubleshooting steps on the Model 10 RFC press...

Page 24: ...t still displays proceed with the following 2 Attempt gentle reseating of the NovaSure disposable device A Partially retract the array into the sheath by releasing the disposable device handle lock re...

Page 25: ...ot appear Model 10 RFC Remove the NovaSure disposable device from the uterus after fully retracting the disposable device array into the sheath A Release the disposable device lock release button B Ho...

Page 26: ...ent into an outlet on a circuit different from that to which the other device s is are connected Contact Hologic Technical Support or the manufacturer of the other equipment for assistance 5 The contr...

Page 27: ...t s hand placed on her thigh should be avoided for example by insertion of dry gauze between skin surfaces WARNING With any electrosurgical device the potential for arcing exists and neuromuscular sti...

Page 28: ...lectromagnetic site survey should be considered If the measured field strength in the location in which the RF generator is used exceeds the applicable RF compliance level above the RF generator shoul...

Page 29: ...non Hologic Manufactured Equipment is warranted through its manufacturer and such manufacturer s warranties shall extend to Hologic s customers to the extent permitted by the manufacturer of such non...

Page 30: ...Follow instructions for use Fragile Fuse High pressure HP Humidity limitation non condensing Keep dry Manufacturer No oil Not made with natural rubber latex Protect from heat Radio frequency RF energ...

Page 31: ...MAN 03523 001 Rev 009...

Reviews: