Hologic NovaSure Advanced Instructions For Use And Controller Operator'S Manual Download Page 5

4

ENGLISH

BACK TO TOC

• If the cavity integrity assessment fails after reasonable attempts 

to implement the troubleshooting procedures (step 2 .36), abort 
the procedure .

• For patients in whom the procedure was aborted due to a suspected 

uterine wall perforation, a work-up for perforation should be 
considered prior to discharge .

General

• Endometrial ablation using the NovaSure system is not a sterilization 

procedure . Therefore, the patient should be advised of appropriate 
birth control methods .

• Endometrial ablation does not eliminate the potential for endometrial 

hyperplasia or adenocarcinoma of the endometrium and may mask the 
physician’s ability to detect or make a diagnosis of such pathology . 

• Endometrial ablation is intended for use only in women who do 

not desire to bear children because the likelihood of pregnancy is 
significantly decreased following the procedure . Pregnancy following 
ablation may be dangerous for both mother and fetus .

• Patients who undergo endometrial ablation procedures who have 

previously undergone tubal ligation are at increased risk of developing 
post ablation tubal sterilization syndrome which can require 
hysterectomy . This can occur as late as 10 years post procedure .

• A health hazard may exist in the case where the NovaSure procedure 

is performed in the presence of a thermally and electrically conductive 
Essure

®

 micro-insert that is improperly positioned (e .g ., perforating the 

fallopian tube or the myometrium) . If this occurs, heat can be drawn 
away from the intended treatment area toward other tissue and/or 
organs in contact with the conductive object, which may be sufficient 
to cause localized burns . As a result, correct placement of the Essure 
micro-insert must be confirmed, e .g ., by obtaining the report of the 
Essure Confirmation Test (ECT), prior to performing the NovaSure 
procedure . If the report and/or ECT films are unavailable, the ECT 
should be repeated to confirm correct placement of the Essure micro-
inserts prior to performing the NovaSure procedure .

Technical

• Do not use the sterile, single-patient use disposable device if the 

packaging appears to be damaged or there is evidence of tampering .

• The disposable device is for single-patient use only . Do not reuse or 

re-sterilize the disposable device .  

• If any hysteroscopy procedure is performed with hypotonic solution 

immediately prior to NovaSure treatment, then the uterine cavity must 
be flushed with normal saline prior to treatment with the NovaSure 
system . The presence of hypotonic fluid may reduce the efficiency of 
the NovaSure system .

• Plugging the disposable device into the controller starts CO

2

 flow to 

purge any air out of the disposable device and tubing . This purging 
operation takes approximately 10 seconds and 

must be performed 

with the disposable device external to the patient to eliminate 
the risk of air or gas embolism

 . The NovaSure RF controller CAVITY 

ASSESSMENT LED flashes red (Model 08-09 RFCs) or a purging device 

screen appears (Model 10 RFC) and an audible pulsed tone sounds 
throughout the purge procedure . When the tone and the LED/screen 
message stops it is safe to insert the disposable device .

• For patients with cardiac pacemakers or other active implants, a 

possible hazard exists due to interference with the action of the 
pacemaker that may occur and may damage the pacemaker . Consult 
the pacemaker manufacturer for further information when use of the 
NovaSure system is planned in patients with cardiac pacemakers .

• Care should be taken to ensure the patient does not contact metal 

parts which are earthed or which have an appreciable capacitance to 
earth, such as direct contact with the metal on tables .

• 

Danger: explosion hazard . 

Do not use in the presence of a flammable 

anesthetic mixture . Do not use in the presence of flammable gases or 
liquids .

• Failure of the NovaSure  RF controller could result in an unintended 

increase in output power .

PRECAUTIONS

• It has been reported in the literature that patients with a severely 

anteverted, retroflexed or laterally displaced uterus are at greater risk 
of uterine wall perforation during any intrauterine manipulation .

• A false passage can occur during any procedure in which the uterus 

is instrumented, especially in cases of severe anteverted retroflexed 
or a laterally displaced uterus . Use caution to ensure that the device is 
properly positioned in the uterine cavity .

• The NovaSure system consists of the following components:

-  single-patient use NovaSure disposable device with connecting cord
-  NovaSure RF controller
-  NovaSure CO

2

 canister

-  NovaSure desiccant
-  NovaSure foot switch
-  power cord

  To ensure proper operation, never use other components with the 

NovaSure system . Inspect the components regularly for damage, 
and do not use them if damage is apparent . The use of any cables or 
accessories other than those specified in these instructions may result 
in increased emissions or decreased immunity of the RF controller .

• The RF controller must be installed and put into service according 

to the guidance provided in these instructions to ensure its 
electromagnetic compatibility . Refer to the electromagnetic emissions 
and immunity tables in the Specifications section .

• The RF controller should not be used adjacent to or stacked with other 

equipment . If adjacent or stacked use is necessary, the RF controller 
should be observed to verify normal operation in the configuration in 
which it will be used .

• Portable and mobile RF communications equipment can affect the 

RF controller . Refer to the electromagnetic immunity tables in the 
Specifications section for recommended separation distances .

• Patients who have undergone endometrial ablation and are later 

placed on hormone replacement therapy should have a progestin 

Summary of Contents for NovaSure Advanced

Page 1: ......

Page 2: ...d the disposable device may contain a more recent revision of the NovaSure system instructions than the manual provided with the controller The NovaSure disposable device is not to be used with other...

Page 3: ...suctioning The disposable device is connected to the controller via a cord containing the RF cable suction tubing used for pressure monitoring during the cavity integrity assessment cycle and for suct...

Page 4: ...hyperplasia a patient with any anatomic condition e g history of previous classical cesarean section or transmural myomectomy or pathologic condition e g long term medical therapy that could lead to...

Page 5: ...the disposable device external to the patient to eliminate the risk of air or gas embolism The NovaSure RF controller CAVITY ASSESSMENT LED flashes red Model 08 09 RFCs or a purging device screen app...

Page 6: ...after inserting the disposable device and proceed directly from the seating procedure to the CIA Electrically conductive objects e g monitoring electrodes from other devices that are in direct contact...

Page 7: ...e g hormone D C or patient timing Patients randomized into the control arm received wire loop resection as an endometrial pretreatment Study endpoints The primary effectiveness measure was a validated...

Page 8: ...rameters between the treatment groups between the age groupings or among the nine investigational sites Table 2 Patient Accountability Number of Patients NovaSure Wire Loop Resection Plus Rollerball E...

Page 9: ...culty Performing Work Or Other Activities Due to Menses Pre operatively 66 3 65 5 Table 4 Effectiveness Quality of Life QOL NovaSure Wire Loop Resection Plus Rollerball 12 Months 9 9 8 6 24 Months 14...

Page 10: ...use only in women who do not desire to bear children because the likelihood of pregnancy is significantly decreased following the procedure Patients of childbearing capacity should be cautioned of pot...

Page 11: ...5 Press the toggle switch on the back panel of the controller into the on position 1 6 Connect the foot switch to the appropriate port on the front panel of the controller NOTE The first time the Mode...

Page 12: ...ay can be extended CONTRAINDICATION Do not treat a patient with a uterine cavity length that is less than 4 cm as cervical canal damage may occur NOTE Patients with a uterine cavity length greater tha...

Page 13: ...2 18 Unlock the disposable device by pressing the lock release button Close the disposable device by holding the front handle stationary and gently pulling the rear handle backwards until the closed...

Page 14: ...eeze the disposable device handles together while gently moving the disposable device 0 5 cm to and from the fundus and rotating the handle of the disposable device 45 counterclockwise from the vertic...

Page 15: ...m 2 34 Select the value indicated on the WIDTH dial into the NovaSure RF controller input screen by depressing the UP DOWN arrows 2 35 The system can be operated in either automatic mode or manual mod...

Page 16: ...ting 24 Replacement Instructions 24 Specifications 24 Cleaning and Sanitizing 28 Parts List 28 Warranty 28 Technical Support and Product Return Information 28 Symbol Definitions 29 MODEL 08 09 RF CONT...

Page 17: ...o be at the cervix and cannot be resolved by using the cervical collar use another tenaculum to grasp the cervix around the sheath Repeat the CIA by pressing the foot switch NOTE CO2 leakage may occur...

Page 18: ...modification of this equipment is allowed Periodic maintenance Recommended periodic maintenance includes RF power output testing and standard safety tests such as leakage current in accordance with UL...

Page 19: ...ated when the array is not fully deployed The ENABLE LED cannot be toggled on nor can power be delivered to the array when the ARRAY POSITION LED is illuminated VACUUM LED illuminates in two condition...

Page 20: ...e new device abort the procedure NOTE Removing the disposable device from the uterine cavity after completing a cavity integrity assessment will require an additional CIA test to be performed upon dis...

Page 21: ...2 leakage may occur at the external cervical os due to the presence of an over dilated cervix Visible bubbles or the hissing sound of escaping gas may accompany CO2 leakage under either of these condi...

Page 22: ...ng and standard safety tests such as leakage current in accordance with UL 60601 1 along with regular cleanings described in the Cleaning and Sanitizing section RF power output test The NovaSure RF co...

Page 23: ...r dilated cervix Visible bubbles or the hissing sound of escaping gas may accompany CO2 leakage under either of these conditions NOTE When following the troubleshooting steps on the Model 10 RFC press...

Page 24: ...t still displays proceed with the following 2 Attempt gentle reseating of the NovaSure disposable device A Partially retract the array into the sheath by releasing the disposable device handle lock re...

Page 25: ...ot appear Model 10 RFC Remove the NovaSure disposable device from the uterus after fully retracting the disposable device array into the sheath A Release the disposable device lock release button B Ho...

Page 26: ...ent into an outlet on a circuit different from that to which the other device s is are connected Contact Hologic Technical Support or the manufacturer of the other equipment for assistance 5 The contr...

Page 27: ...t s hand placed on her thigh should be avoided for example by insertion of dry gauze between skin surfaces WARNING With any electrosurgical device the potential for arcing exists and neuromuscular sti...

Page 28: ...lectromagnetic site survey should be considered If the measured field strength in the location in which the RF generator is used exceeds the applicable RF compliance level above the RF generator shoul...

Page 29: ...non Hologic Manufactured Equipment is warranted through its manufacturer and such manufacturer s warranties shall extend to Hologic s customers to the extent permitted by the manufacturer of such non...

Page 30: ...Follow instructions for use Fragile Fuse High pressure HP Humidity limitation non condensing Keep dry Manufacturer No oil Not made with natural rubber latex Protect from heat Radio frequency RF energ...

Page 31: ...MAN 03523 001 Rev 009...

Reviews: