background image

English

6

MyoSure

®

 XL Tissue Removal Device for Fluent

Connecting the Tissue Removal Device to the Fluent Fluid Management System

1. Remove the tissue removal device (REF 50-601XL/50-603XL) from the sterile package.
2. Sterile person hands the flexible drive cable and suction tube to the non-sterile person.
3.  Non-sterile person inserts the flexible cable into the corresponding adapter on the Fluent 

Fluid Management System as shown in Figure 3.

4.  The tissue removal device flexible drive cable has a keyed feature that serves to align the 

handpiece cable to the Fluent Fluid Management System connector. The metal tab on the 
connector is pushed down, the flexible cable inserted and then the tab is released.

CAUTION:  

DO NOT attempt to sharply bend the flexible drive cable in a diameter of less 
than 8 inches (20 centimeters). A sharply bent or kinked drive cable may 
cause the Fluent Fluid Management System motor to overheat and stop. During 
a procedure, a minimum distance of 5 feet  (1.5 meters) should be maintained 
between the Fluent Fluid Management System and the tissue removal device to allow the drive cable to hang in a large arc 
with no bends, loops, or kinks.

5.  Non-sterile person attaches the tissue removal device suction tube to the corresponding fitting on the Out-FloPak.

Operation  

1.  Set up the Fluent Fluid Management System for a MyoSure tissue removal procedure per the 

instructions in the Fluent Fluid Management System Operating Manual.

2.  The foot pedal activates tissue removal device operation. The foot pedal turns the motor ON 

and OFF. Once the foot pedal is depressed, the tissue removal device accelerates and rotates 
to the set speed and continues until the foot pedal is released.

3.  Press the foot pedal and observe the tissue removal device action to verify that the motor 

runs and that the cutting window is closed as shown in Figure 4.

WARNING:  

Periodic irrigation of the tissue removal device tip is recommended to provide 
adequate cooling and to prevent accumulation of excised materials in the 
surgical site.

4.  Introduce the tissue removal device through the straight 4 mm working channel of a  

hysteroscope.

5.  Under direct hysteroscopic visualization, position the tissue removal device’s side facing  

cutting window against target pathology.

CAUTION: 

 Excessive leverage on the tissue removal device does not improve cutting performance and, in extreme cases, may result in 
wear, degradation, and seizing of the cutter assembly.

6. Press the foot pedal to activate the tissue removal device’s cutting blade.
7.  The tissue removal device’s reciprocating action alternately opens and closes the device’s cutting window to the outflow pump of the Fluent 

Fluid Management System thereby drawing tissue into the cutting window.

8.  Cutting takes place when the tissue removal device cutting edge rotates and translates through the tissue removal device’s cutting window.

CAUTION:  

If it appears that the blade has stopped rotating during a procedure, check to ensure that the tissue removal device is 
properly connected to the Fluent Fluid Management System, all cables to the Fluent Fluid Management System are secure, 
and that the drive cable has not wrapped into a loop.

NOTE:

  If the Fluent Fluid Management System powers down unexpectedly, leave it off for 15 seconds, restart the system and follow 

on-screen prompts.

Figure 3. Insert drive cable and foot pedal into 

the Fluent Fluid Management System.

Figure 4. Closed Tissue Removal Device 

cutting window on left

Summary of Contents for MyoSure XL

Page 1: ...his document contains proprietary data of Hologic Inc No disclosure reproduction or use of any part thereof may be made except by written permission from Hologic REV RELEASE DATE 15 MAY 2019 Artwork c...

Page 2: ...t you should be experienced in hysteroscopic surgery with powered instruments Healthy uterine tissue can be injured by improper use of the tissue removal device Use every available means to avoid such...

Page 3: ...ain distension If this situation occurs tap the foot pedal once or twice to run the tissue removal device the cutting window will then close automatically If it appears that the tissue removal device...

Page 4: ...ay be affected RF emissions CISPR 11 Class A The FLUENT FLUID MANAGEMENT SYSTEM is suitable for use in all establish ments other than domestic and those directly connected to the public low volt age p...

Page 5: ...ended separation distance calcu lated from the equation applicable to the frequency of the transmitter Recommended safety distance d 1 2 P for 150 KHz to 80 MHz d 1 2 P for 80 MHz to 800 MHz d 2 3 P f...

Page 6: ...jects and people Tissue Removal Device for Fluent 50 601XL 50 603XL The MyoSure XL Tissue Removal Device for Fluent is shown in Figure 1 It is a hand held unit which is connected to the Fluent Fluid M...

Page 7: ...s to the set speed and continues until the foot pedal is released 3 Press the foot pedal and observe the tissue removal device action to verify that the motor runs and that the cutting window is close...

Page 8: ...pressly stated in the Agreement i Equipment manufactured by Hologic is warranted to the original Customer to perform substantially in accordance with published product specifications for one 1 year st...

Page 9: ...ox with the returned unit Return used or opened product according to the instructions provided with the Hologic supplied biohazard kit For More Information For technical support or reorder information...

Page 10: ...xide U S federal law restricts this device to sale by or on the order of a physician Do not resterilize Do not use if package is damaged Hologic MyoSure and associated logos are registered trademarks...

Reviews: